新型国产BR850电子挥发罐监测七氟烷消耗量液态体积的可行性研究
引用本文:吕锐,高静,杨从文,邓毅,陈妍,陈杨,易斌,鲁开智.新型国产BR850电子挥发罐监测七氟烷消耗量液态体积的可行性研究[J].第三军医大学学报,2017,39(22).
作者姓名:吕锐  高静  杨从文  邓毅  陈妍  陈杨  易斌  鲁开智
作者单位:第三军医大学西南医院手术麻醉科,重庆,400038
基金项目:the Key Science and Technology Project of Chongqing (CSTC2014yykfB10005).重庆市重点科技计划项目
摘    要:目的 评价新型国产BR850电子挥发罐测量法监测七氟烷消耗量液态体积的效果.方法 选择本院2016年1月至2017年6月静吸复合全身麻醉下行妇科腔镜手术患者79名,于Boaray700麻醉机上加载国产BR850电子挥发罐.采用测量法测算术中吸人七氟烷的液态体积,并与传统公式法、称重法进行比较,采用线性回归分析两两之间七氟烷消耗量的相关性.结果 公式法、称重法、测量法测得七氟烷消耗量分别为(33.18±14.06)、(33.87±14.41)、(33.89±14.33) mL,三者两两比较差异均无统计学意义(P>0.05).3种方法两两比较七氟烷消耗量的决定系数R2:称重法与公式法为0.977(P<0.01),称重法与测量法为0.992(P <0.01),公式法与测量法为0.970(P<0.01),三者之间均存在正相关.结论 新型国产BR850电子挥发罐监测术中七氟烷消耗量液态体积与称重法、公式法效果相当,可满足我国精准化医疗的需求.

关 键 词:电子挥发罐  全身麻醉  七氟烷  消耗量  液态体积

Feasibility of monitoring liquid volume of sevoflurane consumption using a new model (BR850) of vaporizer with electronic liquid level sensor
LYU Rui,GAO Jing,YANG Congwen,DENG Yi,CHEN Yan,CHEN Yang,YI Bin,LU Kaizhi.Feasibility of monitoring liquid volume of sevoflurane consumption using a new model (BR850) of vaporizer with electronic liquid level sensor[J].Acta Academiae Medicinae Militaris Tertiae,2017,39(22).
Authors:LYU Rui  GAO Jing  YANG Congwen  DENG Yi  CHEN Yan  CHEN Yang  YI Bin  LU Kaizhi
Abstract:Objective To test the performance of a new model (BR850) of vaporizer equipped with an electronic liquid level sensor for monitoring the liquid volume of sevoflurane consumption.Methods Seventy-nine patients undergoing gynecological laparoscopic surgery under general anesthesia at our hospital between January 2016 and June 2017 were enrolled to evaluate the performance of the BR850 vaporizer loaded on a Boaray700 anesthesia machine.The liquid volume of sevoflurane consumption measured using the electronic liquid level sensor was compared with the measurements calculated using an established formula and derived from weight changes of the vaporizer,and the correlations among the measurements were analyzed.Results The mean sevoflurane consumption measured using the formula,by vaporizer weight changes and using the electronic sensor was 33.18 ± 14.06,33.87 ± 14.41,and 33.89 ± 14.33 mL,respectively.The coefficient of determination (R2) of sevoflurane consumption measurements was 0.977 (P < 0.01) between the weighing method and formula method,0.992 (P < 0.01) between the weighing method and the electronic method,and 0.970 (P < 0.01) between the formula method and the electronic method,showing a positive correlation among the measurements using the 3 methods.Conclusion Compared with the weighing method and the formula method,the electronic method using a liquid level sensor in the vaporizer yields comparable results for measuring sevoflurane consumption during general anesthesia.This new vaporizer can well meet the demand of precision medical treatment in China.
Keywords:electronic anesthetic vaporizer  general anesthesia  sevoflurane  liquid volume  sevoflurane consumption