ICARE回弹式眼压计在婴幼儿青光眼的临床应用
引用本文:冯波,张舒心,王文欣,刘雯. ICARE回弹式眼压计在婴幼儿青光眼的临床应用[J]. 中国实用眼科杂志, 2011, 29(8). DOI: 10.3760/cma.j.issn.1006-4443.2011.08.010
作者姓名:冯波  张舒心  王文欣  刘雯
作者单位:1. 100730北京,首都医科大学附属北京同仁医院,北京同仁眼科中心,北京市眼科学与视觉科学重点实验室
2. 北京儿童医院
摘    要:目的 探讨ICARE回弹式眼压计在婴幼儿青光眼患者的适用性及准确性.方法 青光眼患儿11例(22只眼),其中婴幼儿型青光眼8例(16只眼),青光眼伴其它眼部发育异常3例(6只眼),年龄2.7~19.8个月,平均7.3个月.使用ICARE回弹式眼压计及Schiotz压陷式眼压计进行眼压测量.配对t检验比较两种眼压计之间的差异.结果 所有患儿都可在非用药状态下使用ICARE眼压计测量眼压,Schiotz眼压计须在患儿用药后深睡眠状态下测量眼压,共获得34对可靠数值.ICARE眼压计和Schiotz眼压计的测量值分别为(13.95±10.69)mm Hg和(14.60±10.78)mm Hg(平均值±标准差),二者间的差异无统计学意义(P>0.05).结论 ICARE回弹式眼压计是适合于婴幼儿青光眼患者眼压检查的眼压计,其测量结果 与Schiotz压陷式眼压计测量结果 相似.
Abstract:
Objective To investigate the applicability and accuracy of ICARE rebound tonometer in glaucomatous infants and young children. Methods Eleven cases (22 eyes) of glaucomatous children were involved, including 8 cases (16 eyes) of infantile glaucoma and 3 cases (6 eyes) of glaucoma accompanying other maldevelopment, aged 2.7~19.8 months old. ICARE tonometer and Schiotz tonometer were used to measure intraocular pressure (IOP) of the patients. The measurements were compared used pared student-t test. Results Thirty-four pairs of reliable data were obtained. The values of ICARE and Schiotz tonometer were 13.95t 10.69mmHg and 14.60t 10.78mmHg (averaget SD),respectively. The difference had no statistic signification (P >0.05). Conclusions ICARE Rebound tonometer is suitable to glaucomous infants and young children to measure IOP. The measurements are comparable with those of Schiotz tonometer.

关 键 词:眼压  回弹式眼压计  婴幼儿

Clinical application of ICARE rebound tonometer in glaucomatous infants and young children
FENG Bo,ZHANG Shu-xin,WANG Wen-xin,Liu Wen. Clinical application of ICARE rebound tonometer in glaucomatous infants and young children[J]. Chinese Journal of Practical Ophthalmology, 2011, 29(8). DOI: 10.3760/cma.j.issn.1006-4443.2011.08.010
Authors:FENG Bo  ZHANG Shu-xin  WANG Wen-xin  Liu Wen
Abstract:Objective To investigate the applicability and accuracy of ICARE rebound tonometer in glaucomatous infants and young children. Methods Eleven cases (22 eyes) of glaucomatous children were involved, including 8 cases (16 eyes) of infantile glaucoma and 3 cases (6 eyes) of glaucoma accompanying other maldevelopment, aged 2.7~19.8 months old. ICARE tonometer and Schiotz tonometer were used to measure intraocular pressure (IOP) of the patients. The measurements were compared used pared student-t test. Results Thirty-four pairs of reliable data were obtained. The values of ICARE and Schiotz tonometer were 13.95t 10.69mmHg and 14.60t 10.78mmHg (averaget SD),respectively. The difference had no statistic signification (P >0.05). Conclusions ICARE Rebound tonometer is suitable to glaucomous infants and young children to measure IOP. The measurements are comparable with those of Schiotz tonometer.
Keywords:Intraocular pressure  Rebound tonometer  Infants and young children