Quantitative assessment of right ventricular function and pulmonary regurgitation in surgically repaired tetralogy of Fallot using 256-slice CT: comparison with 3-Tesla MRI |
| |
Authors: | Yuzo Yamasaki Michinobu Nagao Kenichiro Yamamura Masato Yonezawa Yoshio Matsuo Satoshi Kawanami Takeshi Kamitani Ko Higuchi Ichiro Sakamoto Yuichi Shiokawa Hidetake Yabuuchi Hiroshi Honda |
| |
Institution: | 1. Department of Clinical Radiology, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan 2. Department of Molecular Imaging & Diagnosis, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan 3. Department of Pediatrics, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan 4. Department of Cardiovascular Medicine, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan 5. Department of Cardiovascular Surgery, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan 6. Department of Health Sciences, Graduate School of Medical Sciences, Kyushu University, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan
|
| |
Abstract: | Objectives To compare 256-slice cardiac computed tomography (CCT) with cardiac magnetic resonance (CMR) imaging to assess right ventricular (RV) function and pulmonary regurgitant fraction (PRF) in patients with repaired tetralogy of Fallot (TOF). Methods Thirty-three consecutive patients with repaired TOF underwent retrospective ECG-gated CCT and 3-Tesla CMR. RV and left ventricular (LV) end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV) and ejection fraction (EF) were measured using CCT and CMR. PRF-CCT (%) was defined as (RVSV???LVSV)/RVSV. PRF-CMR (%) was measured by the phase-contrast method. Repeated measurements were performed to determine intra- and interobserver variability. Results CCT measurements, including PRF, correlated highly with the CMR reference (r?=?0.71–0.96). CCT overestimated RVEDV (mean difference, 17.1?±?2.9 ml), RVESV (12.9?±?2.1 ml) and RVSV (4.2?±?2.0 ml), and underestimated RVEF (?2.6?±?1.0 %) and PRF (?9.1?±?2.0 %) compared with CMR. The limits of agreement between CCT and CMR were in a good range for all measurements. The variability in CCT measurements was lower than those in CMR. The estimated effective radiation dose was 7.6?±?2.6 mSv. Conclusions 256-slice CCT can assess RV function and PRF with relatively low dose radiation exposure in patients with repaired TOF, but overestimates RV volume and underestimates PRF. Key points ? 256-slice CT assessment of RV function is highly reproducible in repaired TOF. ? Pulmonary regurgitation can be evaluated by biventricular systolic volume difference. ? CT overestimates RV volume and underestimates pulmonary regurgitation, compared with MRI. |
| |
Keywords: | |
本文献已被 SpringerLink 等数据库收录! |
|