Development of a compact portable driver for a pneumatic ventricular assist device |
| |
Authors: | Tomohiro Nishinaka MD Yoshiyuki Taenaka MD Eisuke Tatsumi MD Hiroyuki Ohnishi MD Akihiko Homma PhD Kyoko Shioya VMD Toshihide Mizuno VMD Tomonori Tsukiya PhD Sadahiko Mushika MS Yasuhiro Hashiguchi MS Akira Suzuki MS Soichiro Kitamura MD |
| |
Affiliation: | (1) Department of Artificial Organs, National Cardiovascular Center Research Institute, 5-7-1 Fujishirodai, Suita 565-8565, Japan;(2) Senko Medical Instrument Manufacturing, Tokyo, Japan;(3) National Cardiovascular Center, Osaka, Japan |
| |
Abstract: | The Toyobo-National Cardiovascular Center pneumatic ventricular assist device (Toyobo-NCVC VAD) is widely used in Japan; however, the current pneumatic drivers have some drawbacks, including their large size, heavy weight, and high power consumption. These issues cause difficulty with mobility and contribute to an unsatisfactory quality of life for patients. Because it is urgently necessary to improve patients’ safety and quality of life, we have developed a compact, low-noise, portable VAD driver by utilizing an electrohydraulic actuator consisting of a brushless DC motor and a regenerative pump. This unit can be actuated for as long as 2 h with two rechargeable lightweight batteries as well as with external AC power. It is compact in size (33 × 25 × 43 cm) and light in weight (13 kg), and the unit is carried on a mobile wheeled cart. In vitro testing with a Toyobo-NCVC VAD demonstrated a sufficient pumping capacity of up to 8 l/min. We conclude that this newly-developed compact portable driver can provide a better quality of life and improved safety for patients using protracted pneumatic VAD support. |
| |
Keywords: | Artificial heart Ventricular assist device Pneumatic driver |
本文献已被 SpringerLink 等数据库收录! |
|