Comparison of cool tip versus 8-mm tip catheter in achieving electrical isolation of pulmonary veins for long-term control of atrial fibrillation: a prospective randomized pilot study |
| |
Authors: | Dixit Sanjay Gerstenfeld Edward P Callans David J Cooper Joshua M Lin David Russo Andrea M Verdino Ralph J Patel Vickas V Kimmel Stephen E Ratcliffe Sarah J Hsia Henry H Nayak Hemal M Zado Erica Ren Jian-Fang Marchlinski Francis E |
| |
Affiliation: | From the Cardiovascular Division, Hospital of The University of Pennsylvania and Center for Clinical Epidemiology and Biostatistics, University of Pennsylvania, Philadelphia, Pennsylvania, USA |
| |
Abstract: | Objective: To compare safety and efficacy of 8-mm versus cooled tip catheter in achieving electrical isolation (EI) of pulmonary veins (PV) for long-term control of atrial fibrillation (AF). Background: There is paucity of studies comparing safety/efficacy of 8-mm and cooled tip catheters in patients undergoing AF ablation. Methods and Results: This was a randomized and patient-blinded study. Subjects were followed by clinic visits (at 6 weeks and 6 months) and transtelephonic monitoring (3-week duration) done around each visit. Primary endpoints were: (1) long-term AF control (complete freedom and/or >90% reduction in AF burden on or off antiarrhythmic drugs at 6 months after a single ablation), and (2) occurrence of serious adverse events (cardiac tamponade, stroke, LA-esophageal fistula, and/or death). Eighty-two patients (age 56 ± 9 years, 60 males, paroxysmal AF = 59) were randomized (42 patients to 8-mm tip and 40 patients to cooled tip). EI of PVs was achieved in shorter time by the 8-mm tip as compared with cooled tip catheter (40 ± 23 minutes vs 50 ± 30 minutes; P < 0.05) but long-term AF control was not different between the two (32 patients [78%] vs 28 patients [70%], respectively; P = NS). One serious adverse event occurred in each group (LA-esophageal fistula and stroke, respectively) and no significant PV stenosis was observed in either. Conclusion: EI of PVs using either 8-mm or cooled tip catheter results in long-term AF control in the majority after a single ablation procedure, with comparable efficacy and safety. |
| |
Keywords: | atrial fibrillation pulmonary vein catheter ablation |
本文献已被 PubMed 等数据库收录! |
|