首页 | 本学科首页   官方微博 | 高级检索  
     


Minimum 10-year radiographic follow-up of a cementless acetabular component for primary total hip arthroplasty with a bulk autograft
Authors:Shigehiro Inoue  Motoyuki Horii  Hiroshi Suehara  Keiichiro Ueshima  Toshiki Shiga  Mikihiro Fujioka  Kenji Takahashi  Takeshi Asano  Wook-Choel Kim  Masako Nakagawa  Toshikazu Kubo
Affiliation:(1) Department of Orthopaedic Surgery, Kyoto Prefectural University of Medicine, Kawaramachi-Hirokoji, Kamigyo-ku, Kyoto, 602-8566, Japan;(2) School of Nursing, Faculty of Medicine Mie University, Tsu, Japan
Abstract:We radiographically investigated 36 hips in 34 patients with osteoarthritis of the hip who had undergone total hip arthroplasty. Their mean age was 59.2 years (range 36–79 years), and the mean follow-up period was 11.2 years (range 10–14 years). The long-term outcome and the chronological changes in the bulk autograft were examined. The acetabular component of the prosthesis was a Lord-type threaded cup with a smooth surface. At follow-up, bone absorption was minor in 17 joints, moderate in 11, and major in 8. The hips with graft coverage of gE20% (group A) had a significantly higher loosening rate than hips with coverage of Lt20% (group B) (P Lt 0.05). The cup position changed markedly in group A. Our findings indicate that graft coverage should be less than 20% when a bulk graft is used together with a smooth-surfaced cementless cup.
Keywords:Bone graft  Radiography  Loosening  Cementless
本文献已被 ScienceDirect SpringerLink 等数据库收录!