应用Micro-CT对小鼠全身脂肪扫描成像研究
引用本文:侯豹, 黄术兵, 韩佩宇, 姚文茜, 蔡维维, 曲秀霞, 孙海建, 邱丽颖. 应用Micro-CT对小鼠全身脂肪扫描成像研究[J]. 分子影像学杂志, 2024, 47(2): 111-117. doi: 10.12122/j.issn.1674-4500.2024.02.01
作者姓名:侯豹  黄术兵  韩佩宇  姚文茜  蔡维维  曲秀霞  孙海建  邱丽颖
作者单位:江南大学无锡医学院,江苏 无锡 214122
基金项目:国家自然科学基金(82370364);江苏省实验动物协会2023年项目(DWXH202310)
摘    要:目的  探讨扫描电压和X射线滤光片对小鼠全身脂肪扫描图像质量和对小鼠辐射剂量的影响。方法  选取C57BL/6小鼠6只,3只为正常对照小鼠,3只为高脂喂养的肥胖小鼠。应用Micro-CT对小鼠全身进行成像扫描,统计不同扫描条件下对小鼠的辐射剂量;使用Analyze12.0分析软件对小鼠皮下和内脏脂肪进行提取,然后评价扫描图像质量以及脂肪分布。结果  正常小鼠和肥胖小鼠的扫描结果表明,在扫描电压为70 kV和Al 1.0 mm X射线滤光片扫描条件下,小鼠皮下脂肪和内脏脂肪边界清晰且无伪影,图像质量高,辐射剂量小。使用Analyze 12.0软件分离的小鼠皮下脂肪和内脏脂肪连续且平滑,脂肪显示完整。结论  本研究为临床研究患者皮下脂肪和内脏脂肪提供了理论依据,对CT扫描电压和X射线滤光片的选择提供了数据参考。在降低扫描电压和保证图像质量的前提下,最大限度的降低患者受到的辐射剂量。

关 键 词:Micro-CT   扫描电压   滤光片   脂肪   图像质量
收稿时间:2023-11-27

Application of Micro-CT in whole-body fat scanning imaging of mice
HOU Bao, HUANG Shubing, HAN Peiyu, YAO Wenxi, CAI Weiwei, QU Xiuxia, SUN Haijian, QIU Liying. Application of Micro-CT in whole-body fat scanning imaging of mice[J]. Journal of Molecular Imaging, 2024, 47(2): 111-117. doi: 10.12122/j.issn.1674-4500.2024.02.01
Authors:HOU Bao  HUANG Shubing  HAN Peiyu  YAO Wenxi  CAI Weiwei  QU Xiuxia  SUN Haijian  QIU Liying
Affiliation:Wuxi Medical College, Jiangnan University, Wuxi 214122, China
Abstract:Objective To investigate the effects of scanning voltage and X-ray filter on image quality and radiation dose of whole body fat scanning in mice. Methods Six C57BL/6 mice were selected, including three were normal control mice and three obese mice fed with high fat diet. Micro-CT was used to scan the whole body of mice, and the radiation dose under different scanning conditions was calculated. The subcutaneous and visceral fat were extracted by Analyze12.0 software, and the image quality and fat distribution were evaluated. Results The scanning results of normal mice and obese mice showed that under the scanning voltage of 70 kV and Al 1.0 mm X-ray filter, the boundary between mice subcutaneous fat and visceral fat was clear and without artifacts, with high image quality and low radiation dose. The subcutaneous fat and visceral fat of mice separated by analyze12.0 software were continuous and smooth with complete fat display. Conclusion This study provides a theoretical basis for the clinical study of subcutaneous fat and visceral fat in patients, and provides data reference for the selection of CT scan voltage and filter. Under the premise of reducing the scanning voltage and ensuring the image quality, the radiation dose to the patient was reduced to the maximum.
Keywords:micro-CT  tube voltage  filter  fat  image quality