Prediction of Human Femoral Bone Strength Using Ultrasound Velocity and BMD: An In Vitro Study
Authors:CF Njeh  CW Kuo  CM Langton  HI Atrah  CM Boivin
Institution:(1) Bone Densitometry Service, Queen Elizabeth Hospital, Birmingham, GB;(2) Applied Physics Group, Birmingham University, Birmingham, GB;(3) Centre for Metabolic Bone Disease, Hull Royal Infirmary, Hull, GB;(4) Bone Bank, West Midland Blood Transfusion Service, Birmingham, UK, GB
Abstract:The stiffness and strength of cancellous bone are important in the management of skeletal diseases such as osteoporosis. These properties are a function not only of bone density but also of bone architecture, some measure of which can be provided by quantitative ultrasound. The ability of quantitative ultrasound and bone mineral density (BMD) to predict stiffness and strength of human femoral heads removed from live subjects during hip replacement was assessed. Stiffness and strength were measured using a uniaxial compression test. Ultrasound velocity was measured using the pulse-submersion technique (McClue CUBAResearch) and BMD using DXA (Lunar DPX-L). Ultrasound velocity (quantitative ultrasound) and stiffness varied with the three orthogonal directions, the highest significance being between the proximo-distal (PD) and antero-posterior (AP) directions (p < 0.0001) for stiffness and p = 0.0003 for velocity). Ultrasound velocity was significantly correlated with compressive bone strength (r = 0.76, p < 0.0001) and stiffness (r = 0.79–0.83, p < 0.0001). BMD was also significantly correlated with compressive strength (r = 0.82, p < 0.0001) and stiffness (r = 0.66–0.81, p < 0.001). Using multiple regression analysis both BMD and velocity were significant predictors of strength (r = 0.88, p = 0.0004 and 0.0054 respectively) and stiffness r = 0.92, p = 0.0001 and 0.0003 respectively). BMD and velocity were still independent significant predictors of both stiffness (r = 0.93, p < 0.0001 and 0.0001 respectively) and strength (r = 0.89, p < 0.0001 and 0.02) when they combined as a product (BMDn*Vm). This suggests that BMD measured using DXA, if used in conjunction with ultrasound velocity, may be able to improve osteoporosis risk assessment. The information about femur anisotropy may also be important for hip prosthesis and in vivo modelling.
Keywords:: Cancellous bone  DXA  Osteoporosis  Stiffness  Structure  Ultrasound (QUS)
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号