首页 | 本学科首页   官方微博 | 高级检索  
检索        


Implant survival of the most common cemented total hip devices from the Nordic Arthroplasty Register Association database
Authors:Mika Junnila  Inari Laaksonen  Antti Eskelinen  Pekka Pulkkinen  Leif Ivar Havelin  Ove Furnes
Institution:1. Department of Orthopaedics and Traumatology, Turku University Hospital, Turku;2. Coxa Hospital for Joint Replacement, Tampere;3. The Finnish Arthroplasty Register;4. Department of Public Health, Helsinki University, Helsinki, Finland;5. The Norwegian Arthroplasty Register, Department of Orthopaedic Surgery, Haukeland University Hospital, Bergen;6. Department of Clinical Medicine, University of Bergen, Bergen, Norway
Abstract:Background and purpose — According to previous Nordic Arthroplasty Register Association (NARA) data, the 10-year implant survival of cemented total hip arthroplasties (THAs) is 94% in patients aged 65–74 and 96% in patients aged 75 or more. Here we report a brand-level comparison of cemented THA based on the NARA database, which has not been done previously.

Patients and methods — We determined the rate of implant survival of the 9 most common cemented THAs in the NARA database. We used Kaplan-Meier analysis with 95% CI to study implant survival at 10 and 15 years, and Cox multiple regression to assess survival and hazard ratios (HRs), with revision for any reason as endpoint and with adjustment for age, sex, diagnosis, and femoral head material.

Results — Spectron EF THA (89.9% (CI: 89.3–90.5)) and Elite THA (89.8% (CI: 89.0–90.6)) had the lowest 10-year survivorship. Lubinus (95.7% survival, CI: 95.5–95.9), MS 30 (96.6%, CI: 95.8–97.4), and C-stem THA (95.8%, CI: 94.8–96.8) had a 10-year survivorship of at least 95%. Lubinus (revision risk (RR)?=?0.77, CI: 0.73–0.81), Müller (RR =0.83, CI: 0.70–0.99), MS-30 (RR =0.73, CI: 0.63–0.86), C-stem (RR =0.70, CI: 0.55–0.90), and Exeter Duration THA (RR =0.84, CI: 0.77–0.90) had a lower risk of revision than Charnley THA, the reference implant.

Interpretation — The Spectron EF THA and the Elite THA had a lower implant survival than the Charnley, Exeter, and Lubinus THAs. Implant survival of the Müller, MS 30, CPT, and C-stem THAs was above the acceptable limit for 10-year survival.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏