Efficacy of single lead VDD pacing in patients with impaired and normal left ventricular function
Authors:Schuchert A  Jakob M  Treese N  Schöpfel A  Schmidt W  Jung W  Kreuzer J  Staedt U  Stertmann W A  Meinertz T
Institution:Department of Cardiology, University-Hospital Eppendorf, Hamburg, Germany.
Abstract:Atrial synchronous ventricular pacing seems to be the best pacing mode for patients with advanced AV block and impaired LV function. The long-term follow-up of single lead VDD pacing was studied in 33 patients with impaired LV function and compared to 42 patients with normal LV function. All patients received the same VDD lead and VDDR pacemaker. The lead model with 13-cm AV spacing between the atrial and ventricular electrode was implanted in 89% of the patients. Follow-ups were 1, 3, 6, and 12 months after implantation. The percentage of atrial sensing and the P wave amplitude were determined at each follow-up. Minimal P wave amplitude at implantation was 2.0 +/- 1.4 mV in patients with impaired and 1.7 +/- 0.9 mV with normal LV function (not significant). At the 12-month follow-up, 33 patients with normal and 23 patients with depressed LV function remained paced in the VDD mode. The remaining patients died in five (impaired LV function) and seven cases (normal LV function) or their pacemakers were programmed to the VVI/VVIR pacing mode in four (impaired LV function) and three cases (normal LV function). P wave amplitude did not differ in the two groups (e.g., at month 12: impaired: 1.17 +/- 0.42 mV; normal: 1.09 +/- 0.49 mV). The atrial sensitivity was programmed in most patients to sensitive settings with no differences between the two groups (e.g., at month 12: impaired: 0.13 +/- 0.06 mV; normal: 0.13 +/- 0.05 mV). The diagnostic counters indicated nearly permanent atrial sensing (e.g., at month 12: impaired: 99.3 +/- 2.2%; normal: 99.0 +/- 1.0 mV). In conclusions, single lead VDD pacing restored AV synchronous ventricular pacing in patients with normal and with impaired LV function indicating that it could be an alternative to DDD pacemakers, but not to dual-chamber pacing.
Keywords:single lead VDD pacing  VDD pacemaker  left ventricular function  advanced AV block  P wave amplitude
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号