首页 | 本学科首页   官方微博 | 高级检索  
     


Tendon transfer fixation: comparing a tendon to tendon technique vs. bioabsorbable interference-fit screw fixation
Authors:Sabonghy Eric Peter  Wood Robert Michael  Ambrose Catherine Glauber  McGarvey William Christopher  Clanton Thomas Oscar
Affiliation:Department of Orthopedics, The University of Texas Medical School at Houston, Houston, TX 77030, USA. psabonghy@hotmail.com
Abstract:Tendon transfer techniques in the foot and ankle are used for tendon ruptures, deformities, and instabilities. This fresh cadaver study compares the tendon fixation strength in 10 paired specimens by performing a tendon to tendon fixation technique or using 7 x 20-25 mm bioabsorbable interference-fit screw tendon fixation technique. Load at failure of the tendon to tendon fixation method averaged 279N (Standard Deviation 81N) and the bioabsorbable screw 148N (Standard Deviation 72N) [p = 0.0008]. Bioabsorbable interference-fit screws in these specimens show decreased fixation strength relative to the traditional fixation technique. However, the mean bioabsorbable screw fixation strength of 148N provides physiologic strength at the tendon-bone interface.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏