Effect of Inner Taper Angle of Acetabular Metal Shell on the Malseating and Dissociation Force of Ceramic Liner
Authors:Young-Kyun Lee  Ki-Chul Kim  Woo-Lam Jo  Yong-Chan Ha  Javad Parvizi  Kyung-Hoi Koo
Affiliation:1. Department of Orthopedic Surgery, Seoul National University Bundang Hospital, Seongnam, Republic of Korea;2. Department of Orthopedic Surgery, Dankook University Hospital, Cheonan, Chungcheongnam-do, Republic of Korea;3. Department of Orthopedic Surgery, Catholic University of Korea School of Medicine, Seoul, Seoul, Republic of Korea;4. Department of Orthopedic Surgery, Chung-Ang University College of Medicine, Seoul, Republic of Korea;5. Department of Orthopedic Surgery, Rothman Institute, Thomas Jefferson University, Philadelphia, Pennsylvania;6. Department of Orthopedic Surgery, Seoul National University College of Medicine, Seoul, Republic of Korea
Abstract:

Background

Malseating of ceramic liner appeared as a matter of concern in multibearing metal, although multibearing cup can be coupled with hard liners as well as polyethylene liner. In this metal shell, the inner taper angle should be 10° for the modularity, while standard metal shells for ceramic liner have an inner taper angle of 18°. However, there has been no study in the effect of taper angle to the risk of malseating. We evaluated whether the taper angle of metal shell might affect the malseating of ceramic liner, and dissociation force of ceramic liner from metal shell.

Methods

Three surgeons manually inserted ceramic liners into two designs of metal shell with different tapers angles (10° and 18°). We compared malseating rates of ceramic liners and push-out strengths, which means dissociation force of the ceramic liner from the metal shell, between these two metal shell designs.

Results

The malseating rates in 10° metal shell were higher than those in 18° metal shell (23.3% vs 0%, P < .05). The mean dissociation force (1148.8 ± 46.7 N) in 10° taper cup was higher than that (389.7 ± 108.3 N) in 18° taper cup (P < .01).

Conclusion

Our results suggest that surgeon should be cautious about malseating of ceramic liner when using multi-bearing metal shell with inner taper angle of 10°.

Clinical relevance

When surgeon use multi-bearing metal shell with inner taper angle of 10°, our results suggest that surgeon should be cautious about malseating of ceramic liner.
Keywords:total hip arthroplasty  ceramic liner  compatibility  malseating  push-out force