Evaluation of implant abutment screw tightening protocols on reverse tightening values: An in vitro study
Authors:Abdullah H. Alnasser  Chandur P.K. Wadhwani  Todd R. Schoenbaum  Mathew T. Kattadiyil
Affiliation:1. Graduate Program student in Advanced Education Program in Prosthodontics, Loma Linda University, Loma Linda, Calif;1. Teaching Assistant, Prosthodontics Department, School of Dentistry, Prince Sattam Bin Abdulaziz University, Al-Kharj, Saudi Arabia;2. Adjunct Assistant Professor, Advanced Education Program in Prosthodontics, Loma Linda University, Loma Linda, Calif;3. Associate Clinical Professor, Continuing Dental Education, University of California-Los Angeles, Los Angeles, Calif;4. Professor and Director, Advanced Education Program in Prosthodontics, Loma Linda University, Loma Linda, Calif
Abstract:Statement of problemImplant abutment screw loosening is a common prosthetic complication of implant-supported crowns. However, reports that have objectively evaluated the effectiveness of different tightening protocols on reverse tightening values are sparse.PurposeThe purpose of this in vitro study was to determine the optimal tightening protocol for implant abutment screws.Material and methodsFifty Neoss implants were randomly distributed to 5 groups (n=10). The implants received a cover screw and mounted, and the impression coping was tightened. Tightening was measured by using a digital measuring device. Then, the implant abutments were placed and tightened to 32 Ncm by using a Crystaloc screw. In Group 2T10I, the screws were tightened twice with an interval of 10 minutes between the first and second tightening. In Group 2T0I, the screws were tightened twice with no interval time. In Group 1T, the screws were tightened 1 time only. In Group TCT, the screws were tightened, counter-tightened, and then tightened again. In Group TCTCT, the abutment screws were tightened, counter-tightened, tightened, counter-tightened, and then tightened again. All the mounted implants were left in the same environment for 3 hours, and the reverse tightening values were then measured.ResultsThe mean reverse tightening values of the first 4 groups ranged from 21.49 Ncm to 22.57 Ncm, whereas the reverse tightening value for the fifth group was 25.51 Ncm. A significant difference was found among the groups (P<.05) with reverse tightening data.ConclusionsNo significant difference was found in tightening the abutment screw 2 times with a 10-minute interval time, no interval time, or tightening it 1 time only. However, a significant difference was found in reverse tightening in the 3-time tightening and counter-tightening group.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号