3种图像模式在影像引导放射治疗中的差异性研究
引用本文:胡彩容,张秀春,陆军,蔡勇君,吴君心,潘建基.3种图像模式在影像引导放射治疗中的差异性研究[J].中华放射医学与防护杂志,2012,32(4):374-378.
作者姓名:胡彩容  张秀春  陆军  蔡勇君  吴君心  潘建基
作者单位:福建省肿瘤医院放疗科,福州,350014
基金项目:福建省卫生厅青年科研课题(2011-1-23)
摘    要:目的 检测并分析电子射野影像系统( EPID)、kV级平面摄像(kV planar)、kV级锥形束CT( CBCT)3种图像模式在影像引导放疗(IGRT)中摆位误差的精度.方法 通过在计划系统的模体CT图像上模拟25组摆位偏差,生成25组DRR参考图像和25组三维CT参考图像,分别与模体在原点位置获取的EPID,kV planar和CBCT 3种图像模式进行图像配准和摆位误差的测量,检测并比较3种图像模式在摆位误差测量中的精度.结果 共读取675组残留误差,3种图像及其相应的配准方法均具有较高的精确度,残留误差的平均值在x、y、z方向上均<1 mm,CBCT的灰度值配准精度<0.1 mm,kV planar图像在手动配准时其测量摆位误差的精度好于EPID(<0.65 mm),在自动配准方法时与EPID具有同等的精度.结论 医院IGRT系统的3种图像模式及其相应的配准方法均可以满足临床应用,CBCT为首选,结合图像质量、成像剂量和配准精度等因素,kV planar优于EPID.

关 键 词:影像引导  图像模式  残留误差  配准精度
收稿时间:2011/10/24 0:00:00

A comparision of three imaging modalities in image-guided radiotherapy
HU Cai-rong,ZHANG Xiu-chun,LU Jun,CAI Yong-jun,WU Jun-xin and PAN Jian-ji.A comparision of three imaging modalities in image-guided radiotherapy[J].Chinese Journal of Radiological Medicine and Protection,2012,32(4):374-378.
Authors:HU Cai-rong  ZHANG Xiu-chun  LU Jun  CAI Yong-jun  WU Jun-xin and PAN Jian-ji
Institution:Department of Radiation Oncology, the Teaching Hospital of Fujian Medical University, Fujian Provincial Tumor Hospital, Fuzhou 350014, China;Department of Radiation Oncology, the Teaching Hospital of Fujian Medical University, Fujian Provincial Tumor Hospital, Fuzhou 350014, China;Department of Radiation Oncology, the Teaching Hospital of Fujian Medical University, Fujian Provincial Tumor Hospital, Fuzhou 350014, China;Department of Radiation Oncology, the Teaching Hospital of Fujian Medical University, Fujian Provincial Tumor Hospital, Fuzhou 350014, China;Department of Radiation Oncology, the Teaching Hospital of Fujian Medical University, Fujian Provincial Tumor Hospital, Fuzhou 350014, China;Department of Radiation Oncology, the Teaching Hospital of Fujian Medical University, Fujian Provincial Tumor Hospital, Fuzhou 350014, China
Abstract:Objective To investigate and analyze the positioning accuracy of three imaging modalities utilized in image-guided radiotherapy (IGRT): electronic portal imaging device (EPID), kV portal image (kV planar) and the kV cone beam computed tomography (CBCT).Methods 25 groups of setup errors were simulated on the phantom images through treatment planning system. Digitally reconstructed radiographs (DRRs) were constructed from the CT data which were subsequently used as references to register the EPID and kV planar images acquired at the original position. In addition, the reconstructed 3D-CT images were used to register the CBCT images. Finally, the setup errors using several registration methods were measured to investigate and compare the accuracies of the three imaging modalities used for patient setup. Results 675 groups of residual errors were analyzed. All combinations of imaging modalities and registration method were found to be accurate. The mean residual errors in three directions were less than 1 mm. The method based on grey value match of CBCT images was found as the most accurate with an uncertainty below 0.1 mm. When the manual match was used, the performance of kV planar was more accurate than that of EPID (residual error <0.65 mm). If automatic registration was applied, kV planar generated similar results as EPID did. Conclusions The three available imaging modalities and their corresponding registration methods are all competent for the clinical application of IGRT in our department. Considering the image quality, radiation dose and the accuracy of registration, CBCT has the priority on IGRT followed by the kV planar.
Keywords:Image-guided radiotherapy  Imaging modality  Residual error  Registration accuracy