Reliability of force/displacement measures in a clinical device designed to measure ligamentous laxity at the knee
Authors:Emery M  Rodgers M M  Boerman J  Fleming B C  Howe J  Pope M
Abstract:This research was supported by a grant from The Foundation for Physical Therapy, Alexandria, VA. The purpose of this study was to determine the reliability of a quantitative measure of knee ligament laxity in vivo. The testing device simultaneously measures applied forces and resulting displacement at the knee through use of a multidimensional electrogoniometer, dynamometer, and a computer-assisted analysis system. Subjects were 30 females and 20 males, ranging in age from 18-32 years. Laxity tests were performed in the anterior/posterior (A/P) plane at 90 and 30 degrees of knee flexion, and the varus/valgus (V/V) plane at 20 and 0 degrees of knee flexion. Test-retest (N = 41) and intertester (N = 18) repeatability was assessed in all test positions by ANOVA and by coefficients of correlation and variability. No significant differences were found within testers for A/P or V/V tests 1 week apart at any angle tested. There were significant differences between testers for A/P and V/V tests. No significant differences were found between left and right sides, or between male/female values. Comparisons to other in vitro and in vivo studies are made. J Orthop Sports Phys Ther 1989;10(11):441-447.