Performance des glottiscopes : une étude comparative randomisée menée sur simulateur d’intubation difficile
Authors:J Sudrial  W Abdi  R Amathieu  D Luis  P Jabre  B Le Roux  V Slavov  W Kamoun  A Avenel  J-L Sebbah  X Combes  G Dhonneur
Institution:1. Département d’anesthésie-réanimation, CHU Jean-Verdier, avenu du 14-Juillet, 93140 Bondy, France;2. Smur 95-Urgences, CHG Gonesse, 27, rue Pierre-de-Theilley, BP 30071, 95503 Gonesse, France;3. Samu 94, CHU Henri-Mondor, 51, avenue du Maréchal-de-Lattre-de-Tassigny, 94000 Créteil, France
Abstract:

Introduction

We have prospectively compared simulated-difficult tracheal intubation characteristics of four glottiscopes: Airtraq™, GlideScope™, McGrath™, LMA CTrach™ with that of the conventional Macintosh laryngoscope.

Study design

prospective with the airway devices proposed in a randomly assigned order.

Materials and method

Forty-two physicians, naïve to glottiscope handling accepted participating this study after the learning curve of each airway device was completed. Participants were requested to perform two series of five tracheal intubations on the manikin Airman™, the first in standard situation and the second in difficult tracheal intubation simulation. The airway devices were chosen in a randomly assigned order. For each airway tool, the following tracheal intubation characteristics were recorded: laryngeal exposure quality, tracheal intubation and apnea durations. A performance index was calculated and a tracheal intubation difficulty was measured during simulation.

Results

More than 1600 supervised tracheal intubations were performed, including 1000 for the learning process of the glottiscopes, which was completed after 10 uses on the manikin. During standard situation, laryngeal exposure quality was similar with the five airway devices. As compared to the Macintosh laryngoscope, GlideScope™, McGrath™, tracheal intubation duration was shorter (p < 0.05) with the Airtraq™ and longer (p < 0.01) with the LMA CTrach™. During difficult tracheal intubation simulation, laryngeal exposure and tracheal intubation duration was of better quality and shorter with the four glottiscopes as compared to that of LM, respectively. Performance index during difficult tracheal intubation simulation simulation was significantly more important (p < 0.01) with the Airtraq™ and the LMA CTrach™. Airtraq™ and Macintosh laryngoscope were respectively the simplest (p < 0.01) and the most difficult (p < 0.01) airway devices to manage a simulated difficult tracheal intubation.

Conclusion

When difficult airway was simulated on the manikin, the four glottiscopes were superior to the Macintosh laryngoscope to improve laryngeal exposure quality and to reduce duration of tracheal intubation. Airtraq™ and the LMA CTrach™ both demonstrated remarkable advantage over GlideScope™ and McGrath™ for simulated difficult intubation management.
Keywords:Intubation difficile simulé  e  GlideScope®    Airtraq®    McGrath®    LMA CTrach®    Laryngoscope de Macintosh