Comparison of electrical velocimetry and cardiac magnetic resonance imaging for the non-invasive determination of cardiac output
Authors:Frederik Trinkmann  Manuel Berger  Christina Doesch  Theano Papavassiliu  Stefan O Schoenberg  Martin Borggrefe  Jens J Kaden  Joachim Saur
Institution:1.1st Department of Medicine (Cardiology, Angiology, Pneumology, Intensive Care), University Medical Center Mannheim, Medical Faculty Mannheim,Heidelberg University,Mannheim,Germany;2.Institute of Clinical Radiology and Nuclear Medicine, University Medical Center Mannheim, Medical Faculty Mannheim,Heidelberg University,Mannheim,Germany;3.DZHK (German Center for Cardiovascular Research) Partner Site, University Medical Center Mannheim, Medical Faculty Mannheim,Heidelberg University,Mannheim,Germany
Abstract:A novel algorithm of impedance cardiography referred to as electrical velocimetry (EV) has been introduced for non-invasive determination of cardiac output (CO). Previous validation studies yielded diverging results and no comparison with the non-invasive gold standard cardiac magnetic resonance imaging (CMR) has been performed. We therefore aimed to prospectively assess the accuracy and reproducibility of EV compared to CMR. 152 consecutive stable patients undergoing CMR were enrolled. EV measurements were taken twice before or after CMR in supine position and averaged over 20 s (AESCULON®, Osypka Medical, Berlin, Germany). Bland–Altman analysis showed insufficient agreement of EV and CMR with a mean bias of 1.2 ± 1.4 l/min (bias 23 ± 26 %, percentage error 51 %). Reproducibility was high with 0.0 ± 0.3 l/min (bias 0 ± 8 %, percentage error 15 %). Outlier analysis revealed gender, height, CO and stroke volume (SV) by CMR as independent predictors for larger variation. Stratification of COCMR in quintiles demonstrated a good agreement for low values (<4.4 l/min) with bias increasing significantly with quintile as high as 3.1 ± 1.1 l/min (p < 0.001). Reproducibility was not affected (p = 0.71). Subgroup analysis in patients with arrhythmias (p = 0.19), changes in thoracic fluid content (p = 0.51) or left heart failure (p = 0.47) could not detect significant differences in accuracy. EV showed insufficient agreement with CMR and good reproducibility. Gender, height and increasing CO and SV were associated with increased bias while not affecting reproducibility. Therefore, absolute values should not be used interchangeably in clinical routine. EV yet may find its place for clinical application with further investigation on its trending ability pending.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号