Clinical Experience with the Intertach 262-12 Pulse Generator in Patients with Recurrent Supraventricular and Ventricular Tachycardia
Authors:MARTIN FROMER  HANS GLOOR  TERESA KUS  LUKAS KAPPENBERGER  MOHAMMAD SHENASA
Institution:The Division of Cardiology, Centre Hospitalier Universitaire Vaudois (CHUV), Lausanne, Switzerland;*Division of Cardiology, Medical Clinic, University Hospital, Zurich, Switzerland;**Electrophysiology Laboratory, Research Center, Hôpital du Sacré-Coeur de Montreal, University of Montreal, Quebec, Canada
Abstract:FROMER, M., ET AL.: Clinical Experience with the Intertach 262-12 Pulse Generator in Patients with Recurrent Supraventricular and Ventricular Tachycardia. An antitachycardia pulse generator, the Intermedics Intertach 262-12 was implanted in 16 patients (14 patients with supraventricular tachycardia of various origins and two patients with recurrent ventricular tachycardia), who were not responsive to various antiarrhythmic drug regimens. The follow-up was from 6–49 months (mean 30.9 ± 13.8). Five patients had a follow-up of over 3 years. The device was used in all patients. One patient with ventricular tachycardia died from a nonarrhythmic cause. Loss of responsiveness to burst pacing was observed in 1/14 patients with supraventricular tachycardia and nontolerance of antitachycardia pacing in one patient. Overall clinical success of pacing was observed in 13/16 patients = 81%. The pacemaker proved to be a versatile system with reliable tachycardia detection and termination functions.
Keywords:antitachycardia pacing  supraventricular tachycardia  ventricular tachycardia