Experimental Studies With a 9F Forward‐Looking Intracardiac Imaging and Ablation Catheter
Authors:Douglas N. Stephens MS  Matthew O'Donnell PhD  Kai Thomenius PhD  Aaron Dentinger PhD  Douglas Wildes PhD  Peter Chen PhD  K. Kirk Shung PhD  Jonathan Cannata PhD  Pierre Khuri-Yakub PhD  Omer Oralkan PhD  Aman Mahajan MD  PhD  Kalyanam Shivkumar MD  PhD  David J. Sahn MD
Affiliation:1. University of California, Davis, California USA;2. University of Washington, Seattle, Washington USA;3. GE Global Research, Schenectady, New York USA;4. Irvine Biomedical Inc/St Jude Medical, Irvine, California USA;5. University of Southern California, Los Angeles, California USA;6. Stanford University, Palo Alto, California USA;7. University of California, Los Angeles, California USA;8. Oregon Health & Science University, Portland, Oregon USA
Abstract:Objective. The purpose of this study was to develop a high‐resolution, near‐field‐optimized 14‐MHz, 24‐element broad‐bandwidth forward‐looking array for integration on a steerable 9F electrophysiology (EP) catheter. Methods. Several generations of prototype imaging catheters with bidirectional steering, termed microlinear (ML), were built and tested as integrated catheter designs with EP sensing electrodes near the tip. The wide‐bandwidth ultrasound array was mounted on the very tip, equipped with an aperture of only 1.2 by 1.58 mm. The array pulse echo performance was fully simulated, and its construction offered shielding from ablation noise. Both ex vivo and in vivo imaging with a porcine animal model were performed. Results. The array pulse echo performance was concordant with Krimholtz‐Leedom‐Matthaei model simulation. Three generations of prototype devices were tested in the right atrium and ventricle in 4 acute pig studies for the following characteristics: (1) image quality, (2) anatomic identification, (3) visualization of other catheter devices, and (4) for a mechanism for stabilization when imaging ablation. The ML catheter is capable of both low‐artifact ablation imaging on a standard clinical imaging system and high–frame rate myocardial wall strain rate imaging for detecting changes in cardiac mechanics associated with ablation. Conclusions. The imaging resolution performance of this very small array device, together with its penetration beyond 2 cm, is excellent considering its very small array aperture. The forward‐looking intracardiac catheter has been adapted to work easily on an existing commercial imaging platform with very minor software modifications.
Keywords:ablation  electrophysiology  interventional guidance  intracardiac ultrasound array  miniaturized
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号