首页 | 本学科首页   官方微博 | 高级检索  
     


28 mm Metasul™ bearings in primary total hip arthroplasty
Authors:C. Delaunay
Affiliation:(1) Department of orthopaedic surgery, clinique de I’Yvette, 67-71, route de Corbeil, F-91160 Longjumeau, France
Abstract:Assessment of possible low-wear with some former metal-on-metal (MoM) total hip arthroplasties (THA) led to the reintroduction of metallic bearings in the late 80’s. The author reports on two studies of Metasul-28 mm cementless THA. In the first one in a general population, impingement has been the main cause of osteolysis and Co level survey has been a good indicator of Metasul bearing behaviour. In the second study, in a group of 83 less than 50-year-old and active patients, Metasul bearings showed good wear resistance at 7.2 years mean follow-up. In both studies, no general toxic effect could have been detected thus far. According to the current knowledge, it is always reasonable to expect low-wear and better THA longevity with use of MoM bearings under the following conditions: 1) use of a CoCr alloy with high carbide concentration; 2) reduce impingement risk (head without sleeve, slimmer as possible neck, perfectly adapted Morse cone from the same manufacturer, well — oriented components); and 3) prefer cementless acetabular fixation.
Keywords:Total hip arthroplasty  Metal-on-metal bearings  Impingement  Metallic ions