Temperature Dependent Pulse Generator Dysfunction
Authors:ARNE M. ABRAHAMSEN,TORBJØ  RN AARSLAND
Affiliation:Division of Cardiology, Medical Department, Central Hospital of Rogaland, Stavanger, Norway
Abstract:Malfunction of a lithium-powered R-wave inhibited unipolar pulse generator (CPI Model 0505) necessitated removal 15 months after implantation. Examination of the pulse generator in a water bath revealed abnormal function at temperatures above 31.5 degrees C and normal function at temperatures below 31.5 degrees C. CPI suggested that the failure was caused by the 10 kHz timing crystal in the generator.
Keywords:malfunction of pacemaker    programming failure    programmable pacemaker    temperature dependent pacemaker function    timing crystal