Abstract: | Background: The ProSeal(TM) Laryngeal Mask Airway (PLMA) (Laryngeal Mask Company, Henley-on-Thames, United Kingdom) is a new laryngeal mask with a modified cuff designed to improve its seal and a drain tube for gastric tube placement. Similarly, the Laryngeal Tube Suction (LTS) (VBM Medizintechnik Gmbh, Sulz a.N, Germany) is a new laryngeal tube that also has an additional channel for gastric tube placement. This study compared the placement and functions of these two devices. Methods: One hundred fifty patients undergoing general anesthesia for elective surgery were randomly allocated to the PLMA (n = 75) or LTS (n = 75). Oxygenation and ventilation, ease of insertion, fiberoptic view, oropharyngeal leak pressure, ventilatory data, ease of gastric tube insertion, and postoperative airway morbidity were determined. Results: After successful insertion of the devices in 96% of patients with the PLMA and in 94.4% with the LTS it was possible to maintain oxygenation, ventilation, and respiratory mechanics during the entire duration of surgery. Successful first and second attempt insertion rates were 57 patients (76%) and 15 patients (20%), respectively, for the PLMA and 60 patients (80%) and 11 patients (14.6%), respectively, for the LTS. Airway placement was unsuccessful with the PLMA in three patients and with the LTS in four patients. Time to achieve an effective airway was 36 +/- 24 s with the PLMA versus 34 +/- 25 s with the LTS. Gastric tube insertion was possible in 97.3% of patients with the PLMA and in 96% with the LTS. |