Inflammatory system activation during cardiopulmonary bypass as an indicator of biocompatibility: a randomized comparison of bubble and membrane oxygenators
Authors:L Nilsson  U Nilsson  P Venge  O Johansson  H Tydén  T Aberg  S O Nystr?m
Affiliation:Department of Thoracic Surgery, University Hospital, Uppsala, Sweden.
Abstract:As the exposure of blood to foreign material during cardiopulmonary bypass (CPB) leads to triggering of inflammatory systems, the inflammatory response was used as an indicator of the biocompatibility of oxygenators. Activation of complement and neutrophil granulocytes during CPB was studied in 96 patients undergoing coronary bypass, with randomized comparisons between four different oxygenators, two of bubble and two of membrane type. Seven patients undergoing thoracotomy without CPB served as controls. During CPB there was significant complement activation, measured as changes in the ratio C3d/C3, with no demonstrable difference between the bubble and membrane oxygenator groups. Such change was not seen in the controls. Neutrophil granulocytes released significant amounts of the granule proteins lactoferrin and myeloperoxidase during CPB, but not during thoracotomy without CPB. The plasma concentrations of lactoferrin and myeloperoxidase were significantly lower in the membrane oxygenator groups, possibly indicating better biocompatibility. The strong inflammatory response with both oxygenator types, however, indicates that presently used CPB devices have unsatisfactory biocompatibility.