Design of a motion simulation system to assist respiratory gating for radiation therapy
Affiliation:1. Division of Radiation Oncology, Departments of Oncology, National Taiwan University Hospital, Taipei, Taiwan;2. Institute of Electro-Optical Science and Technology, National Taiwan Normal University, Taipei, Taiwan;3. Graduate Institute of Biomedical Electronics and Bioinformatics, National Taiwan University, Taipei 106, Taiwan;4. Department of Radiation Oncology, Fu Jen Catholic University Hospital, New Taipei City, Taiwan;5. School of Medicine, College of Medicine, Fu Jen Catholic University, New Taipei City, Taiwan;6. Graduate Institute of Oncology, National Taiwan University College of Medicine, Taipei, Taiwan
Abstract:Stereotactic ablative radiotherapy (SABR) aims to deliver high doses of radiation to kill cancer cells and shrink tumors in less than or equal to 6 fractions. However, organ motion during treatment is a challenging issue for this kind of technique. We develop a control system via Bluetooth technology to simulate and correct body motion during SABR. Methods: Radiation doses were analyzed, and the radiation damage protection capability was checked by external beam therapy 3 (EBT3) films irradiated by a linear accelerator. A wireless signal test was also performed. A validation was performed with 8 previously treated patient respiratory pattern records and 8 healthy volunteers. Results: The homemade simulation system consisted of 2 linear actuators, one movable stage with a maximal moving distance of 6.5 cm × 12.5 cm × 5 cm to simulate the respiratory pattern of 8 patients precisely with a median error of 0.36 mm and a maximal motion difference of 1.17 mm, and 3.17 and chipset transited signals to display them as a waveform. From the test with 8 volunteers, the chip could detect deep respiratory movement up to 3 cm. The effect of the chip on a radiation dose of 400 monitor units (MUs) by 6 MV photons and 200 MUs by 10 MV photons showed high penetration rates of 98.8% and 98.6%, respectively. Conclusions: We invented a tubeless and wireless respiratory gating detection chip. The chip has minimal interference with the treatment angles, good noise immunity and the capability to easily penetrate a variety of materials. The simulation system consisting of linear actuators also successfully simulates the respiratory pattern of real patients.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号