首页 | 本学科首页   官方微博 | 高级检索  
     


Low-level activity measurement of 131Xem, 133Xem, 135Xe and 133Xe in atmospheric air samples using high-resolution dual X-gamma spectrometry.
Authors:G Le Petit  C Jutier  P Gross  V Greiner
Affiliation:Commissariat à l'Energie Atomique (CEA), DIF/DASE/SRCE, BP 12, 91680 Bruyères-le-Chatel, France. gilbert.le-petit@cea.fr
Abstract:High-resolution dual X and gamma spectrometry has been optimized to improve the sensitivity in analyzing radioxenon from environmental samples, especially for the metastable isotopes (131Xem and 133Xem). The full-energy peak (FEP) efficiency curve in the 20-700 keV range has been established with a global uncertainty better than +/-3% using a Monte Carlo simulation. The minimum detectable activity concentrations (MDC) achievable at the laboratory for 131Xem, 133Xem, 133Xe and 135Xe are presented and the influence of 133Xe activity on the metastable radioxenon MDCs is evaluated.
Keywords:
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号