高频线阵探头与低频凸阵探头在椎动脉检测中的临床研究
引用本文:傅建群,姜惠悦.高频线阵探头与低频凸阵探头在椎动脉检测中的临床研究[J].临床超声医学杂志,2011,13(7):447-450.
作者姓名:傅建群  姜惠悦
作者单位:广东省佛山市南海区第二人民医院超声科,528251
基金项目:佛山市医学类科技攻关项目
摘    要:目的 探讨高频线阵探头(HFLP)与低频凸阵探头(LFCP)在椎动脉检查中的使用价值.方法 分别使用HFLP与LFCP对152例住院患者,共计304支椎动脉进行检查,比较两种探头下椎动脉在颈段(V1)和椎段(V2)的二维图像及彩色多普勒血流显像(CDFI),并将二者检查有问题的血管即阳性结果与磁共振血管造影(MRA)阳...

关 键 词:超声检查  椎动脉  探头  高频率  低频率

Comparative study of high frequency linear-array probe and lower frequency convex probe in detection of vertebral artery
FU Jianqun,JIANG Huiyue.Comparative study of high frequency linear-array probe and lower frequency convex probe in detection of vertebral artery[J].Journal of Ultrasound in Clinical Medicine,2011,13(7):447-450.
Authors:FU Jianqun  JIANG Huiyue
Institution:Department of Ultrasound, the Second People' s Hospital of Nanhai District, Foshan 528251 ,China
Abstract:Objective To observe the application value of high frequency linear probe (HFLP) and low-frequency convex array (LFCP) in examing vertebral artery. Methods A total of 304 VA in 152 patients were examined by HFLP and LFCP, the two -dimensional imaging and color Doppler flow imaging (CDFI)of vertebral artery at Vland V2 segment bewteen the two probes were compared , then the positive results of blood vessels were compared with those of MRA. Results With HFLP and LFCP, the V1 display rate of two - dimensional imaging and CDFI was 0.87 ±0. 34,0.83 ± 0.37 ;0.72 ± 0.45, 0.95±0. 21. The V2 display rate was 3.48 ±0. 66,3.37 ± 0.75;3.19 ± 0.78,4.11 ± 0.62. There was significant difference of display rate between two probes ( P 〈 0.01 ) , there was no significant difference of V1 and V2 display rate between two dimensional imaging and CDFI by HFLP ( P 〉 0.05 ), while there was significant difference by LFCP ( P 〈0. 05 ). Compared with MRA, the positive rate of HFLP and LFCP was 86.02% and 89.25% , respectively, the false positive rate of HFLP and LFCP was 23.81% and 18.63%. Conclusion CDFI can clearly show blood flow signals in the initial segment of V1 and upper V2 with LFCP, which can make up the shortage of HFLP. Routine application of CDFI with LFCP can correct false positive results of HFLP and improve the accuracy of ultrasound in diagnosis of vertebral artery disease.
Keywords:Ultrasonography  Vertebral artery  Probe  high frequency  low frequency