首页 | 本学科首页   官方微博 | 高级检索  
     


Measurement of CT radiation profile width using CR imaging plates
Authors:Liu Ho-Ling  Liu Ruijie Rachel  Reeve Donna M  Shepard S Jeff  Willis Charles E
Affiliation:Department of Imaging Physics, The University of Texas MD Anderson Cancer Center, Houston, Texas 77030, USA.
Abstract:This paper describes the procedure for using a Fuji computed radiography (CR) imaging plate (IP) for the measurement of computed tomography (CT) radiation profiles. Two sources of saturation in the data from the IP, signal and quantization, were characterized to establish appropriate exposure and processing conditions for accurate measurements. The IP generated similar profiles compared to those obtained from digitized ready-pack films, except at the profile edges, where the exposure level is low. However, when IP pixel values are converted to exposure, CR and digitized film profiles are in agreement. The full width at half maximum (FWHM) of the CT radiation profile was determined from the relationship between pixel value and exposure and compared to FWHM of the digitized optical density profile from film. To estimate the effect of scattering by the cassette material, radiation profiles were acquired from IPs enclosed in a cassette or in a paper envelope. The presence of the cassette made no difference in the value determined for FWHM. With proper exposure and processing conditions, the FWHM of 5, 10, and 15 mm collimated beams were measured using IPs to be 7.1, 11.9, and 17.0 mm and using film to be 7.2, 12.2, and 16.8 mm, respectively. Our results suggest that, under appropriate conditions, the estimation of the width of the CT radiation profile using Fuji CR is comparable to the measurement from film density described in American Association of Physicists in Medicine (AAPM) Report No. 39. Although our experiment was conducted using Fuji CR, we anticipate that CR plates from other vendors could be successfully used to measure CT beam profiles because of similar empirical relationships between pixel value and exposure.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号