首页 | 本学科首页   官方微博 | 高级检索  
检索        


Highly Sensitive Detection of Hepatitis B Virus Surface Antigen by Use of a Semiautomated Immune Complex Transfer Chemiluminescence Enzyme Immunoassay
Authors:Kazuhiko Takeda  Mari Maruki  Takahiro Yamagaito  Machiko Muramatsu  Yasuhiro Sakai  Hiroaki Tobimatsu  Hironori Kobayashi  Yoshiteru Mizuno  Yukio Hamaguchi
Institution:Elemental Technology Development Department 1, Technology Development Division, Sysmex Corporation, Kobe, Japan
Abstract:The performance of hepatitis B surface antigen (HBsAg) screening assays is continuously improved to reduce the risk of transfusion-associated hepatitis B. In this study, a semiautomated immune complex transfer chemiluminescence enzyme immunoassay (ICT-CLEIA) for the detection of HBsAg, which is as sensitive as hepatitis B virus (HBV) DNA PCR, was developed; the ICT-CLEIA assay performance was compared with the performance of the Architect HBsAg QT assay and HBV DNA PCR. The specificities in the initial assay and after retesting were 99.50% (1,988/1,998 samples) and 99.95% (1,997/1,998 samples), respectively. The analytical detection limit was determined to be 0.2 mIU/ml using the 2nd International WHO HBsAg standard, and the cutoff value (0.5 mIU/ml) of the ICT-CLEIA assay was 8.0 standard deviations (SD) above the mean of the HBsAg-negative specimens. The ICT-CLEIA assay could detect HBsAg even in the presence of anti-HBs antibodies and demonstrated a 23.6-day-shorter window period using commercially available HBsAg seroconversion panels than the Architect HBsAg QT assay. Furthermore, the monitoring of the viral kinetics by the ICT-CLEIA assay and the HBV DNA PCR produced very similarly shaped curves during both the HBsAg seroconversion and reverse seroconversion periods. Therefore, the ICT-CLEIA assay may be useful not only for an earlier detection of HBV reactivation but also for the monitoring of hepatitis B patients.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号