首页 | 本学科首页   官方微博 | 高级检索  
检索        


Control of radiofrequency lesion size by power regulation
Authors:F H Wittkampf  R N Hauer  E O Robles de Medina
Institution:Department of Cardiology, University Hospital, Utrecht, The Netherlands.
Abstract:The influence of power and exposure duration on lesion size in radiofrequency catheter ablation was investigated in 15 closed-chest dogs. Radiofrequency energy was delivered to the right ventricular endocardium between the tip of a standard 6F electrode catheter and a large W and durations of 5, 10, 20, 30, and 60 seconds. At necropsy 1 week later, well-demarcated homogeneous lesions were found when power had exceeded a threshold level that decreased from 1.8 W at 5 seconds to 0.7 W at 60 seconds. Lesion size ranged from 0 to 7.5 mm in depth and 0 to 9 mm in diameter. For the 5, 10, and 20 second ablations, lesion size was determined by exposure duration and power level. However, after a 20 second exposure, lesion size had reached maturity and was related to delivered power only. Therefore, a gradual, controlled growth of the lesion can be obtained by a stepwise increase of the radiofrequency power level with ample exposure duration at each level to allow for stabilization. At levels exceeding 7 W, the formation of a thin insulating layer of blood coagulum on the electrode surface caused an abrupt increase of impedance within approximately 30 seconds. Therefore, lesion size is limited to 8.5 mm in radiofrequency ablation with a standard 6F endocardial electrode catheter.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏