Biomechanical effect of the configuration of screw hole style on locking plate fixation in proximal humerus fracture with a simulated gap: A finite element analysis
Institution:1. Department of Orthopaedics Tongzhou Luhe Teaching Hospital, Capital Medical University, Beijing, China;2. Department of Physical Therapy and Assistive Technology, National Yang-Ming University, Taipei, Taiwan;3. Department of Electrical Engineering, Chung Yuan Christian University, Taoyuan, Taiwan;4. Department of Biomedical Engineering, Chung Yuan Christian University, Taoyuan, Taiwan;5. Translation Technology Center for Medical Device, Chung Yuan Christian University, Taoyuan, Taiwan;1. Israel National Center for Trauma and Emergency Medicine Research, Gertner Institute for Epidemiology and Health Policy Research, Tel Hashomer, Ramat Gan, Israel;2. Faculty of Medicine, Tel-Aviv University, School of Public Health, Tel-Aviv, Israel;1. Ankara Mevki Military Hospital, Department of Radiology, Ankara, Turkey;2. Bursa Military Hospital, Department of General Surgery, Bursa, Turkey;3. University of Miami Miller School of Medicine, Department of Radiology, Miami, FL, USA;4. Istanbul Arel University, Department of Electrical and Electronics Engineering, Istanbul, Turkey;5. Gulhane Military Medical Academy, School of Medicine, Department of Radiology, Ankara, Turkey;6. Yeni Yuzyil University, Department of Electrical and Electronics Engineering, Istanbul, Turkey;7. Ankara Guven Hospital, Department of General Surgery, Ankara, Turkey;1. Department of Mechanical Engineering, Ben-Gurion University, Beer-Sheva, Israel;2. Department of Mechanical Engineering, Shamoon College of Engineering, Beer-Sheva, Israel;3. Department of Orthopaedics, Hadassah University Hospital, Jerusalem, Israel
Abstract:BackgroundLocking plate fixation for proximal humeral fractures is a commonly used device. Recently, plate breakages were continuously reported that the implants all have a mixture of holes allowing placement of both locking and non-locking screws (so-called combi plates). In commercialized proximal humeral plates, there still are two screw hole styles included “locking and dynamic holes separated” and “locking hole only” configurations. It is important to understand the biomechanical effect of different screw hole style on the stress distribution in bone plate.MethodsFinite element method was employed to conduct a computational investigation. Three proximal humeral plate models with different screw hole configurations were reconstructed depended upon an identical commercialized implant. A three-dimensional model of a humerus was created using process of thresholding based on the grayscale values of the CT scanning of an intact humerus. A “virtual” subcapital osteotomy was performed. Simulations were performed under an increasing axial load. The von Mises stresses around the screw holes of the plate shaft, the construct stiffness and the directional displacement within the fracture gap were calculated for comparison.ResultsThe mean value of the peak von Mises stresses around the screw holes in the plate shaft was the highest for combi hole design while it was smallest for the locking and dynamic holes separated design. The stiffness of the plate-bone construct was 15% higher in the locking screw only design (132.6 N/mm) compared with the combi design (115.0 N/mm), and it was 4% higher than the combi design for the locking and dynamic holes separated design (119.5 N/mm). The displacement within the fracture gap was greatest in the combi hole design, whereas it was smallest for the locking hole only design.ConclusionsThe computed results provide a possible explanation for the breakages of combi plates revealed in clinical reports. The locking and dynamic holes separated design may be a better configuration to reduce the risk of plate fracture.
Keywords:Plate breakage  Combi-hole  Bone fracture  Stress concentration  Finite element method
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号