Validity of glucose measurements in the blood by a glucometer reagent strip in critically ill infants
Authors:Alireza Eskandarifar  Mohammad Aziz Rasouli  Majid Mansouri  Samaneh Moosavi  Ahmad Fotoohi
Affiliation:1. Department of Pediatric, Faculty of Medicine, Kurdistan University of Medical Sciences, Sanandaj, Iran;2. Student Research Committee, Kurdistan University of Medical Sciences, Sanandaj, Iran
Abstract:IntroductionHypoglycemia in critically ill infants is a frequent metabolic disorder which is often due to defective glucose homeostasis. If not recognized annotated early, it can result in severe neurological damage with high mortality and morbidity. Today, glucometer is recommended for fast measuring blood glucose. The aim of this study was to evaluate the accuracy of glucometer for early diagnosis of hypoglycemia in the acutely ill infant.Materials and methodsThis study evaluates 130 critically ill infants less than 1year of age who were admitted to Besat hospital – Sanandaj – IRAN. Blood sugar was measured by standard serum method (glucose oxidase) and glucometer reagent strip.ResultsThe overall means of blood sugar were 115.6 ± 1.75 with serum method and 119.1 ± 1.56 with glucometer method. The rates of sensitivity, specificity, positive predictive value and negative predictive value with glucometer method and with serum glucose method were 72%, 53%, 62%, and 77% respectively. The correlation between the two methods was significant (p < 0.001). Kappa statistics for the two methods was 42%.ConclusionsThe results showed that in general, glucometer may be appropriate for rapid screening in emergency situations and when frequent blood glucose monitoring is needed but it cannot be regarded as a very suitable and reliable tool for diagnosis of hypoglycemia in critically ill infants.
Keywords:Corresponding author.  Glucose  Glucometer  Infant  PICU
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号