首页 | 本学科首页   官方微博 | 高级检索  
     


Concept of the homemade coronary stent: Experimental results and initial clinical experience
Authors:Ivan K. De Scheerder  Kai Wang  Pitsanu Kerdsinchai  Eric Verbeken  Walter Desmet  Jo Dens  Krzysztof Wilczek  Jan Van Humbeeck  Jan H. Piessens
Abstract:To defray the escalating cost of coronary stenting, we handmade a balloon expandable coil stent with stainless steel wire. Preliminary comparison with the Palmaz-Schatz stent showed that, when implanted in porcine illac arteries, there was no difference in immediate angiographic results or in the degree of foreign body reaction at 6 wk. Subsequently, a total of 73 stents were implanted in 52 patients, either as a bailout device (54%) or for suboptimal angiographic results (46%). All but two implantations were successful. The postprocedural regimen consisted of heparin 1,000 IU/hr, aspirin 250 mg daily, and ticlopidine 500 mg daily. In-hospital complications were limited to two groin hematomas, one necessitating blood transfusion. Importantly, stent thrombosis was not observed. While 6-mo follow-up is pending, we already conclude that a balloon expandable coil stent can be handmade easily at low cost and implanted safely in patients. © 1996 Wiley-Liss, Inc.
Keywords:coronary artery disease  coronary angioplasty  coronary stents