实施QC对影像质量与病人剂量影响的研究
引用本文:程玉玺,周启甫,葛丽娟,侯长松,齐雪松,岳保荣,王作岭,尉可道. 实施QC对影像质量与病人剂量影响的研究[J]. 中华放射医学与防护杂志, 1999, 19(6): 400-402
作者姓名:程玉玺  周启甫  葛丽娟  侯长松  齐雪松  岳保荣  王作岭  尉可道
作者单位:100088,北京,卫生部工业卫生实验所
基金项目:国际原子能机构科研基金
摘    要:目的 探讨实施放射诊断设备质量控制(QC) 后,病人受照剂量及X 射线胶片影像质量的变化。方法 QC前后用热释光剂量探测器测量病人入射体表剂量,并用欧洲共同体(CEC) 影像质量标准对已测量剂量的病人X 射线胶片影像质量进行评价,做统计学分析。结果 X 射线机实施质量控制后,病人受照剂量有了明显下降(P< 0-01 或P< 0-05),胸部后前位下降了23% ,腰椎前后位和腰椎侧位下降了48% ,骨盆前后位下降了88% 。QC前、后比较,腰椎侧位片影像质量没有显著差异(P>0-05),而胸部后前位、腰椎前后位及骨盆前后位胶片影像质量反而有所提高( P<0-01 或P<0-05)。结论 对X射线机进行QC检测后可使病人受照剂量较大幅度地下降,但X 射线胶片影像质量没有降低,诊断信息没有丢失。

关 键 词:放射诊断  质量控制  体表入射剂量  影像质量
收稿时间:1998-07-17
修稿时间:1998-07-17

Effect of quality control implementation on image quality of radiographic films and irradiation doses to patients
CHENG Yuxi,ZHOU Qipu,GE Lijuan. Effect of quality control implementation on image quality of radiographic films and irradiation doses to patients[J]. Chinese Journal of Radiological Medicine and Protection, 1999, 19(6): 400-402
Authors:CHENG Yuxi  ZHOU Qipu  GE Lijuan
Affiliation:CHENG Yuxi,ZHOU Qipu,GE Lijuan,et al.Laboratory of Industrial Hygiene,Ministry of Health,Beijing 100088,China,
Abstract:Objective To study the changes in the image quality of radiographic films and the irradiation doses to patients after quality control (QC) implementation. Methods The entrance surface doses (ESD) to patients measured with TLD and the image quality of radiographic films were evaluated on the basis of CEC image quality criteria. Results The ESD to patients were significantly reduced after QC implementation ( P <0 01 or P <0 05)23% for chest PA,48% for lumbar spine AP and LAT,88% for pelvis AP.In comparison with pre QC,the post QC image quality of lumbar spine LAT was not significantly changed ( P >0 05),but the post QC image quality was significantly improved in chest PA,lumbar spine AP and pelvis AP(P0 01 or P <0 05). Conclusion Significantly reduced irradiation dose with improved image quality can be obtained by QC implementation.;
Keywords:Diagnostic radiology Quality control Entrance surface dose Image quality