Accurate Measurement of Agatston Score Using kVp-Independent Reconstruction Algorithm for Ultra-High-Pitch Sn150 kVp CT |
| |
Authors: | Xi Hu Xinwei Tao Yueqiao Zhang Zhongfeng Niu Yong Zhang Thomas Allmendinger Yu Kuang Bin Chen |
| |
Affiliation: | 1.Department of Radiology, Sir Run Run Shaw Hospital, Zhejiang University School of Medicine, Hangzhou, China.;2.Siemens Healthineers China, Shanghai, China.;3.Computed Tomography-Research & Development, Siemens Healthcare GmbH, Erlangen, Germany.;4.Medical Physics Program, University of Nevada, Las Vegas, NV, USA. |
| |
Abstract: | ObjectiveTo investigate the accuracy of the Agatston score obtained with the ultra-high-pitch (UHP) acquisition mode using tin-filter spectral shaping (Sn150 kVp) and a kVp-independent reconstruction algorithm to reduce the radiation dose.Materials and MethodsThis prospective study included 114 patients (mean ± standard deviation, 60.3 ± 9.8 years; 74 male) who underwent a standard 120 kVp scan and an additional UHP Sn150 kVp scan for coronary artery calcification scoring (CACS). These two datasets were reconstructed using a standard reconstruction algorithm (120 kVp + Qr36d, protocol A; Sn150 kVp + Qr36d, protocol B). In addition, the Sn150 kVp dataset was reconstructed using a kVp-independent reconstruction algorithm (Sn150 kVp + Sa36d, protocol C). The Agatston scores for protocols A and B, as well as protocols A and C, were compared. The agreement between the scores was assessed using the intraclass correlation coefficient (ICC) and the Bland–Altman plot. The radiation doses for the 120 kVp and UHP Sn150 kVp acquisition modes were also compared.ResultsNo significant difference was observed in the Agatston score for protocols A (median, 63.05; interquartile range [IQR], 0–232.28) and C (median, 60.25; IQR, 0–195.20) (p = 0.060). The mean difference in the Agatston score for protocols A and C was relatively small (−7.82) and with the limits of agreement from −65.20 to 49.56 (ICC = 0.997). The Agatston score for protocol B (median, 34.85; IQR, 0–120.73) was significantly underestimated compared with that for protocol A (p < 0.001). The UHP Sn150 kVp mode facilitated an effective radiation dose reduction by approximately 30% (0.58 vs. 0.82 mSv, p < 0.001) from that associated with the standard 120 kVp mode.ConclusionThe Agatston scores for CACS with the UHP Sn150 kVp mode with a kVp-independent reconstruction algorithm and the standard 120 kVp demonstrated excellent agreement with a small mean difference and narrow agreement limits. The UHP Sn150 kVp mode allowed a significant reduction in the radiation dose. |
| |
Keywords: | Dual-source CT Coronary artery calcium Agatston score Radiation dose |
|
|