Tissue compensators with use of vinyl lead sheets for head and neck portals on 4-MV x rays |
| |
Authors: | S K Jani E C Pennington |
| |
Affiliation: | Division of Radiation Oncology, University of Iowa College of Medicine, Iowa City 52242. |
| |
Abstract: | Tissue compensators made with a high atomic number material are compact and retain the skin-sparing effect of megavoltage beams. In this note, we have described the use of a commercially available vinyl lead as a material for tissue compensators for use with 4-MV beams on a Clinac-4/80. The dosimetric data obtained included the sheet's lead equivalence, thickness ratios, surface dose and buildup region, and transmission factors. The presence of a compensator did not alter the beam's skin-sparing effects. It was concluded that the vinyl lead sheets allowed an easy and rapid fabrication of a tissue compensator for head and neck portals of up to 15 X 15 cm. |
| |