64排VCT XT前瞻性心电门控低剂量技术对冠状动脉成像的研究
引用本文:张维权,初国新,黄兴,吴祥才,张丹术,陈志仁.64排VCT XT前瞻性心电门控低剂量技术对冠状动脉成像的研究[J].吉林医学,2009,30(21):2556-2558.
作者姓名:张维权  初国新  黄兴  吴祥才  张丹术  陈志仁
作者单位:吉林省人民医院放射科,吉林长春130021
摘    要:目的:评价64排VCTXT前瞻性心电门控低剂量技术对冠状动脉成像的能力及图像质量分析。方法:连续选取152例行冠状动脉CTA检查的患者(A组平均体重指数25.1,心率〈65次/min)行前门控扫描,根据患者的自动毫安量选择管电流及管电压。同时比较回归性心电门控螺旋扫描的30例患者(B组平均体重指数25.7,心率〈72次/min)的剂量,并对两组图像质量进行评价(配对t检验,P〈0.005为差异有统计学意义)。结果:A组前瞻性门控横断面扫描冠状动脉成像射线剂量平均为(2.75±0.95)mSv,所获图像质量完全达到临床诊断需要,B组回顾性心电门控螺旋扫描冠状动脉成像平均剂量为(18.2±2.9)mSv。B组与A组比较差异有统计学意义,射线剂量减少了85.2%。结论:64排VCTXT前瞻性心电门控技术横断面扫描能满足临床诊断需要,同时也非常有效的降低了辐射剂量。

关 键 词:64排VCTXT  心电门控低剂量技术  冠状动脉成像

Study of dose reduction using prospective electrocardiography-triggered coronary scan with 64-slice VCT XT
Institution:ZHANG Wei -quan, CHU Guo -xin ,HUANG Xing , et al (Jilin Provincial People's Hospital, Changehun 130021, China )
Abstract:Objective To compare radiation dose and image quality between prospective electocardiograph-triggered axial scan and retrospective ECG-gatechelical scan in coronary 64-slice CT angiography(CTA).Method Total 152 consecutive patients:group A:Average body mass index(BMI):25.1,heart rate< 65 bpm underwent 64-slice CTA.Using prospective ECG-triggered axial scan protal(120kV automA preference);group B:Average body mass index(BMI):25.7,heart rate<72 bpm.The image quality was blindly evaluated between group A and group B(paired test was used and P<0.05 as a significant difference).Results There was significant difference(t=18.5,P<0.01) between radiation dose of group A(2.75±0.95) mSv] and that of group B (18.2±2.9) mSv],decreased by 85.2%.The image quality of group A could fulfill clinical diagnosis needs.The average radiation dose of using helical scan without optimization was 17.29 mSv which was similar to that of group B(18.13 mSv).Conclusion Prospetive ECG-trigered axial scan in 64-slice coronary CTA can significantly reduce radiation exposure and the image quality can fulfill clinical diagnosis needs.
Keywords:Tomography  X-ray computed  Coronary vessels  Radiation dosage