Implant failure after externally magnetic-controlled nail lengthening
Institution:Department of Orthopedics and Trauma Surgery, Medical University of Vienna, Währinger Gürtel 18-20, Vienna 1090, Austria
Abstract:The number of intramedullary procedures to treat leg length discrepancy (LLD) has risen in the past decade and this has led to increased complications such as nail breakage. The aim of this study was to reveal our experience with implant failure after externally magnetic-controlled telescopic bone lengthening. We observed nail-breakage in 4 out of 24 patients (16.7%) after 6 to 15 months (median 13.5 months) following lower limb lengthening for LLD due to non-union or insufficient bone healing. Three femora and one tibia were treated in one posttraumatic and three congenital cases. Three out of the four patients had an elevated Charleston Comorbidity Index (CCI). Revision surgery involved removal of the broken nail and implantation of a trauma nail. At final follow-up all patients showed subjective satisfaction, achievement of desired lengthening with complete bone healing. Based on our results no unequivocal risk factor including number of previous surgeries, comorbidities and age was identified. In cases of delayed bone formation, early revision surgery with an exchange to a trauma nail has to be considered to prevent breakage of the intramedullary telescopic device.