Hepatic radiofrequency ablation: in vivo and ex vivo comparisons of 15-gauge (G) and 17-G internally cooled electrodes
Authors:K D Song  M W Lee  H J Park  D I Cha  T W Kang  J Lee  J Y Moon  H Rhim
Affiliation:1.Department of Radiology and Center for Imaging Science, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea;2.Department of Radiology, Chung-Ang University Hospital, Seoul, Republic of Korea
Abstract:

Objective:

To compare the performance of the 15-G internally cooled electrode with that of the conventional 17-G internally cooled electrode.

Methods:

A total of 40 (20 for each electrode) and 20 ablation zones (10 for each electrode) were made in extracted bovine livers and in in vivo porcine livers, respectively. Technical parameters, three dimensions [long-axis diameter (Dl), vertical-axis diameter (Dv) and short-axis diameter (Ds)], volume and the circularity (Ds/Dl) of the ablation zone were compared.

Results:

The total delivered energy was higher in the 15-G group than in the 17-G group in both ex vivo and in vivo studies (8.78 ± 1.06 vs 7.70 ± 0.98 kcal, p = 0.033; 11.20 ± 1.13 vs 8.49 ± 0.35 kcal, p = 0.001, respectively). The three dimensions of the ablation zone had a tendency to be larger in the 15-G group than in the 17-G group in both studies. The ablation volume was larger in the 15-G group than in the 17-G group in both ex vivo and in vivo studies (29.61 ± 7.10 vs 23.86 ± 3.82 cm3, p = 0.015; 10.26 ± 2.28 vs 7.79 ± 1.68 cm3, p = 0.028, respectively). The circularity of ablation zone was not significantly different in both the studies.

Conclusion:

The size of ablation zone was larger in the 15-G internally cooled electrode than in the 17-G electrode in both ex vivo and in vivo studies.

Advances in knowledge:

Radiofrequency ablation of hepatic tumours using 15-G electrode is useful to create larger ablation zones.Radiofrequency ablation (RFA) is the most widely used local ablation technique for the management of primary and metastatic liver tumours. However, previous studies have reported that RFA showed a relatively higher local tumour progression rate than did hepatic resection.1,2 One of the most important factors affecting local tumour progression was insufficient tumour-free ablation margin of hepatic parenchyma around the tumour margin.36Several strategies have been developed to obtain sufficient ablation margin. In the aspect of RFA techniques, overlapping technique and combined treatment with transcatheter arterial chemoembolization can be used.79 Another strategy is to use switching monopolar, bipolar or multipolar modes to deliver radiofrequency (RF) energy more efficiently.10,11 Sufficient ablation margin can also be achieved by more efficient electrodes: internally cooled electrode increases the size of ablation zone by preventing charring around the electrode tip.12,13 Perfusion electrodes can also enlarge the ablation zone by increasing electrical conductance and thermal conductivity.1416The diameter of an electrode is also known to be associated with the size of the ablation zone. Theoretically, as the diameter of an electrode becomes larger, the contact surface of the electrode with the surrounding tissue becomes bigger, thereby increasing the active electric field.17,18 As a result, an electrode with a larger diameter is likely to create a larger ablation zone. In a previous study, Goldberg et al17 reported that the extent of coagulation necrosis by RFA increases as the diameter of an electrode increases through an in vivo experimental study. However, this study was performed with an electrode without an internal cooling system. Recently, a clinical study comparing therapeutic efficacy and safety between 15-G and 17-G internally cooled electrodes of RFA for hepatocellular carcinoma was published.19 According to that study, the 15-G internally cooled electrode created a larger ablation volume than did the 17-G electrode. However, the study was limited by selection bias owing to the retrospective study design. In addition, the ablation protocol was not exactly the same between the two groups. Therefore, the issue whether an internally cooled electrode with a larger diameter creates a larger ablation volume should be verified with ex vivo and in vivo experimental studies.The purpose of this experimental study was to compare the performance of the 15-G internally cooled RF electrode with that of the conventional 17-G electrode in both ex vivo and in vivo studies.
Keywords: