首页 | 本学科首页   官方微博 | 高级检索  
     


Important considerations in measurement of attractive force on metallic implants in MR imagers
Authors:Kagetsu, NJ   Litt, AW
Affiliation:Department of Radiology, New York University Medical Center, NY 10016.
Abstract:Almost all reported measurements of the attractive force exerted on metallic implants in magnetic fields of magnetic resonance imagers have been obtained at the portal. This study was done to determine whether the maximum force on an implant is at the portal. Horizontal attractive forces on needles were measured along the axis of the bore at various distances from the portal of a 0.3-T vertical-field magnet and 0.5- and 1.5-T solenoid magnets. Upward vertical forces along the top of the bore were also measured. The horizontal forces 40 cm in from the portals of the 0.5- and 1.5-T magnets were 70%-80% greater than the horizontal forces at the portal. The upward vertical force at the top of the 0.3-T magnet bore, 63 cm in from the portal, was 20 times the horizontal force at the portal and was twice the maximum force measured in the 0.5-T magnet. The maximum force was not at the portal.
Keywords:
点击此处可从《Radiology》浏览原始摘要信息
点击此处可从《Radiology》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号