Pulmonary artery catheterization: a modified technique |
| |
Authors: | McLellan GL; Scalapino MC |
| |
Institution: | Department of Radiology, University Hospital of Jacksonville, University of Florida Health Science Center 32209. |
| |
Abstract: | A modified technique for catheterization of the pulmonary artery was developed. It involves the passage of a tapered, movable-core, J-tipped guide wire across the right ventricle into the pulmonary artery followed by the advancement of a straightened Grollman pigtail catheter. The technique was successful in 34 of 34 pulmonary artery catheterizations. The method avoids prolonged catheter manipulation within the right ventricle. In addition, since the catheter does not cross the tricuspid valve until the guide wire has been advanced, the occasional complication of the pigtail "hooking" on a tricuspid valve leaflet or chordae tendineae during catheter withdrawal and manipulation is prevented. |
| |
Keywords: | |
|
| 点击此处可从《Radiology》浏览原始摘要信息 |
| 点击此处可从《Radiology》下载免费的PDF全文 |