Abstract: | Hepatitis B surface antigen (HBsAg) quantification has been steadily gaining interest as a clinical marker of therapeutic efficacy, for which two commercial assays are currently available: Architect HBsAg QT (Architect) and Elecsys HBsAg II (Elecsys). HBsAg quantification was evaluated using both assays in 126 human immunodeficiency virus (HIV) and hepatitis B virus (HBV)-coinfected patients initiating treatment with tenofovir dipivoxil fumarate. Linear regression and correlation were used to establish the relationship between the two methods. Bland-Altman analysis was performed to determine mean between-assay difference and limits of agreement (LOA) (±2 standard deviations [SD]) both overall and stratified on HBV (hepatitis B envelope antigen [HBeAg] status, replication, genotype, HBV mutants) or HIV (CD4+ cell count) cofactors. There was a significant correlation between Elecsys and Architect assays (correlation coefficient, r = 0.959; P < 0.001). HBsAg quantification using the Elecsys assay was on average 0.200 log10 IU/ml (LOA, −0.500, 0.800) higher than that using Architect, which was consistent across levels of CD4+ cell count, presence of precore and YMDD mutations, and HBeAg status. A slightly larger mean between-assay difference was observed with genotypes A and G (0.196 and 0.201, respectively) versus HBV genotypes D and E (0.036 and 0.030, respectively). Mutations on the S region at position s120/s145 were the only determinant in which the mean between-assay difference in HBsAg quantification was lower than the null value (−0.078). In conclusion, the Elecsys assay, with automatic on-board dilution, is capable of quantifying serum HBsAg levels in HIV-HBV-coinfected patients, with very high correlation with the Architect assay. |