Dual Energy X-Ray Absorptiometry in Small Rats with Low Bone Mineral Density
Authors:M. M. Petersen  P. K. Nielsen  E. Lewin  K. Olgaard
Affiliation:(1) Department of Orthopaedic Surgery, Rigshospitalet, University of Copenhagen, Copenhagen, Denmark, DK;(2) Department of Nephrology, Rigshospitalet, University of Copenhagen, Copenhagen, Denmark, DK
Abstract:The feasibility of dual energy X-ray absorptiometry (DXA) using the Norland XR-26 Mark II bone densitometer for measurements of bone mineral content (BMC) and bone mineral density (BMD) in small rats was evaluated. Thirty-two young, isogenic, Lewis rats (weights from 119 g to 227 g) were used; normal rats (n = 7) and rats with low BMD obtained from three different vitamin D-depleted models (n = 25). DXA measurements were performed using the special software for small animals. Duplicate scans of excised femurs performed at 2 mm/second (pixel size of 0.5 mm × 0.5 mm) were very precise measurements with a coefficient of variation (CV) below 1.6% in animals with normal BMD; in rats with low BMD, the CV was significantly higher (P= 0.02–0.04), 7.8% and 4.4% for BMC and BMD, respectively. Regression analysis demonstrated that these measurements were related to the ash weight (R2 > 98.6%). The CV for measurements of the lumbar spine at 10 mm/second (pixel size 0.5 mm × 0.5 mm) was 2.6% and 2.2% for BMC and BMD, respectively in rats with normal BMD, and again higher (P= 0.03–0.14) in rats with low BMD, 7.3% and 4.7%, respectively, for BMC and BMD. Even though low CVs were obtained for total body duplicate scans (scan speed of 20 mm/second and a pixel size of 1.5 mm × 1.5 mm), the measurements were problematic for accuracy because of an overestimation of both BMC and the area of bone. Using these scan parameters the measurements of total body bone mineral could not be recommended in small rats with low BMD. Received: 21 May 1999 / Accepted: 3 August 2000 / Online publication: 22 December 2000
Keywords:: Small rats —   Bone mineral density —   Bone mineral content —   Dual energy X-ray absorptiometry —   Methodological evaluation
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号