首页 | 本学科首页   官方微博 | 高级检索  
     


Tissue ablation with excimer laser and multiple fiber catheters: effects of optical fiber density and fluence
Authors:Kvasnicka J  Nakamura F  Lange F  Geschwind H J
Affiliation:Unite d'Hemodynamique et de Cardiologie Interventionnelle, Service d'Explorations Fonctionnelles, INSERM, France.
Abstract:This study was performed to assess the ablational properties of 1.4-, 1.7-, and 2.0-mm wire-guided multiple fiber catheters coupled to a XeCl excimer laser. Samples of postmortem human aorta were irradiated in blood at fluences of 40, 50, and 60 mJ/mm 2. Our results indicate that: 1) an increase in the active irradiation coverage results in an increase in the ablation efficiency; 2) the ablation efficiency is not fluence related using the 1.4- or 1.7-mm multiple fiber catheter but efficiency is fluence dependent using the 2.0-mm multiple fiber catheter; 3) the depth of tissue ablated with a multiple fiber catheter depends primarily upon the proportion of the active irradiation coverage at the catheter tip; 4) the 2.0-mm multiple fiber catheter induces craters surrounded by a larger zone of tissue damage than that observed with the 1.4- or 1.7-mm multiple fiber catheter; and 5) the 2.0-mm multiple fiber catheter should be used cautiously for laser angioplasty because of its high penetration and its risk of arterial wall damage.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号