首页 | 本学科首页   官方微博 | 高级检索  
     


Evaluation of a new percutaneous dilatational tracheostomy set apparatus
Authors:Bewsher M S  Adams A M  Clarke C W  McConachie I  Kelly D R
Affiliation:The Intensive Care Unit, Blackpool Victoria Hospital NHS Trust, Whinney Heys Road, Blackpool FY3 8NR, UK.
Abstract:Percutaneous tracheostomy is a well established technique used primarily to assist weaning from mechanical ventilation on many intensive care units. We report our experiences of a total of 36 procedures performed with the new Blue Rhino Percutaneous Tracheostomy Introducer Set developed by Ciaglia. The technique was successful in all cases and was simpler and quicker to perform than with the earlier Ciaglia percutaneous tracheostomy set. Difficulties were encountered when using Shiley tracheostomy tubes. Significant complications included one posterior wall tear and one tracheal cartilage ring fracture.
Keywords:Tracheostomy: percutaneous   dilatational   complications