基于惯性测量单元的全髋关节置换术中髋关节活动度智能测量系统的体外验证
引用本文:郭盛杰,周一新,唐浩,麦宝钧,朱斌杰,陈平,王志华. 基于惯性测量单元的全髋关节置换术中髋关节活动度智能测量系统的体外验证[J]. 中国骨与关节损伤杂志, 2020, 0(4): 350-352
作者姓名:郭盛杰  周一新  唐浩  麦宝钧  朱斌杰  陈平  王志华
作者单位:北京积水潭医院矫形骨科;北京易迈科技有限公司;清华大学
基金项目:北京市属医院科研培育计划项目(PX2018016)。
摘    要:目的开展一项体外验证研究,评估基于惯性测量单元的智能测量系统(IMUHST)在全髋关节置换术中测量髋关节活动度的准确性。方法IMUHST系统由股骨头测量元件及电脑终端组成。采用不透X线的人工骨右侧髋臼磨锉后置入直径60 mm的臼杯,股骨髓腔内置入股骨柄,模拟全髋关节置换术,将IMUHST系统直径36 mm的试模股骨头装配于股骨侧股骨柄上。将CT图像数据测量的髋关节活动度作为参照标准,分析IMUHST系统测量结果(观察组)与参照标准(对照组)之间的差异,并计算组内相关系数(ICC)值。结果观察组与对照组在髋关节屈伸、内收外展、内旋外旋轴线上测量的活动度差异无统计学意义(P>0.05),ICC值均为0.982。结论体外研究证实IMUHST系统测量全髋关节置换术中髋关节活动度具有较高的准确性,该智能辅助系统具有广阔的临床应用前景。

关 键 词:全髋关节置换术  髋关节活动度  惯性测量单元  智能测量系统

In vitro verification of intelligent measurement system for joint range of motion in total hip arthroplasty based on inertia measurement unit sensor
GUSheng-jie,ZHOU Yi-xin,TANG Hao,MAI Bao-jun,ZHU Bin-jie,CHEN Ping,WANG Zhi-hua. In vitro verification of intelligent measurement system for joint range of motion in total hip arthroplasty based on inertia measurement unit sensor[J]. Chinese Journal of Bone and Joint Injury, 2020, 0(4): 350-352
Authors:GUSheng-jie  ZHOU Yi-xin  TANG Hao  MAI Bao-jun  ZHU Bin-jie  CHEN Ping  WANG Zhi-hua
Affiliation:(Department of Orthopedics,Beijing Jishuitcm Hospital,Beijing 100035,China)
Abstract:Objective To evaluate the accuracy of measuring the range of motion of the hip joint using a novel inertia measurement unit based hip smart trial system(IMUHST)during total hip arthroplasty(THA)surgeries.Methods IMUHST system was composed of a smart trial head containing multiple electronic sensors and a terminal computer.A phantom pelvis model was used with the right side hip joint reamed to the diameter of 60 mm and a trial cup was inserted.A phantom femur model was broached and a trial femoral stem was inserted to simulate the THA.A 36 mm diameter smart femoral trial head was mounted to the stem and reduced.Hip range of motion was measured by both the IMUHST and CT scan method.The differences between the IMUHST(study group)and the CT scan method(control group)were analyzed,and intra-class correlation coefficient(ICC)was calculated.Results There was no significant difference between the study group and the control group in hip flexion and extension,adduction and abduction,and internal-external rotation(P>0.05),and the ICC value was 0.982.Conclusion The current in vitro study showed that the IMUHST is accurate in measuring hip ROM,and this system is a promising tool in clinical application.
Keywords:Total hip arthroplasty  Hip range of motion  Inertia measurement unit  Intelligent measurement system