首页 | 本学科首页   官方微博 | 高级检索  
检索        


Wedge factor dependence with depth, field size, and nominal distance--a general computational rule
Authors:Popescu A  Lai K  Singer K  Phillips M
Institution:Department of Radiation Oncology, University of Washington Medical Center, Seattle 98195-6043, USA.
Abstract:We have investigated the dependence of the wedge factors with field size, depth, nominal, and extended distances for 4, 6, 18, and 24 MV photon beams. Analysis of the experimental data suggests a general linear dependence of the wedge factors with field size and depth. The study shows that changes in wedge factors are insignificant (< or = +/-1.0%) with respect to measurements at nominal SSD, SAD, or extended SSD. This independence of the wedge factors on source-to-surface distance was studied for different photon energies (4-24 MV) and for different attenuating wedges (external and internal wedges). For clinical applications, an algorithm is presented to calculate the wedge factor dependence with field size and depth. The new algorithm has been successfully implemented to replace wedge look-up tables for dose and MU calculations in PRISM 1.2 treatment planning system used in our department.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏