首页 | 本学科首页   官方微博 | 高级检索  
     


The Corneal EyeSys System: accuracy analysis and reproducibility of first-generation prototype
Authors:D D Koch  G N Foulks  C T Moran  J S Wakil
Affiliation:Cullen Eye Institute, Baylor College of Medicine, Houston, TX 77030.
Abstract:The EyeSyS Corneal Topography System manufactured by EyeSys Laboratories, Inc., Houston, Tex utilizes a unique data-acquisition design to perform rapid, cost-effective quantitative photokeratoscopy. We evaluated a first-generation prototype system. Repeated testing with this prototype demonstrated an accuracy and reproducibility to within 0.25 D on spherical surfaces. Similar results were obtained in testing aspherical surfaces without refocusing. Data from preliminary clinical studies are displayed for the following corneal conditions: normal, keratoconus, post-keratoplasty, and post-keratoplasty with wound ectasia. Advantages of the EyeSys System include rapid data analysis and display, ease of operation, and compact table-top size.
Keywords:
设为首页 | 免责声明 | 关于勤云 | 加入收藏