首页 | 本学科首页   官方微博 | 高级检索  
检索        


Assessment of stapes mobility by use of a newly developed piezoelectric ceramic device. A preliminary experiment in dogs
Authors:Gyo K  Sato H  Yumoto E  Yanagihara N
Institution:Department of Otolaryngology, Ehime University School of Medicine, Japan.
Abstract:A piezoelectric device was developed for assessment of stapes mobility during middle ear surgery. The device comprises a pair of ceramic bimorph elements: one for activation of the stapes and the other to pick up the vibration as an electric output, which varies in accordance with the stapes mobility, ie, the inverse of the cochlear input impedance (Zsc). The device is compact and easily manipulated even in the narrow surgical field of the ear. However, the measuring range is restricted to between 1 and 10 kHz. Measurement of Zsc was conducted with this device in 5 ears of 5 dogs. The mean magnitude of Zsc increased with frequency in the range from 1 to 10 kHz: 0.95 megohm at 1 kHz and 8.8 megohms at 10 kHz. After fixation of the stapes with dental cement, the magnitude increased to more than 10 megohms, except at 1 kHz. The results suggest that the device is useful in detecting decreases in stapes mobility in patients with chronic otitis media.
Keywords:
本文献已被 PubMed 等数据库收录!