Authors:Hyuk Choi  Seung Hoon Paik  Kwang Ho Lee  Byoung Goo Min  Yong Soon Won
Affiliation:(1) Interdisciplinary Program in Biomedical Engineering Major, Seoul National University, Seoul, South Korea;(2) Department of Biomedical Engineering, College of Medicine, Seoul National University, Seoul, South Korea;(3) Institute of Medical and Biological Engineering, Medical Research Center, Seoul National University, Seoul, South Korea;(4) Korea Artificial Organ Center, Seoul, South Korea;(5) NewheartBio. Co., Ltd., Kyonggi-Do, South Korea;(6) Department of Thoracic and Cardiovascular Surgery College of Medicine, Sooncheonhyang University, 1174 Jung-dong, Wonmi-gu, Bucheon-Si, Gyeonggi-do, 420-767 Seoul, South Korea;(7) School of Mechanical and Aerospace Engineering, Seoul National University, Seoul, South Korea
Abstract:Twin Pulse Life Support, T-PLS™ has received the CE mark (2003) and Korea Food and Drug Administration (KFDA) approval (2004) for short-term application as an Extracorporeal Life Support system (ECLS). T-PLS’s original intention was to apply for not only short-term but also long-term application such as Extracorporeal ventricular assist device (VAD). Hence, a long-term durability test was conducted. The 1-year reliability of the systems tested in this study did not meet the STS/ASAIO standard of 80% reliability with 60% confidence for a 1-year mission life. However, without the disposable units, which are only designed to operate for 6 h, the 1-year reliability exceeded the STS/ASAIO standard of 80% reliability with 60% confidence. In this study, by using the existing analysis methods and analyzing the root cause of the failure used by a numerical analysis. As eliminating or mitigating of the root cause of the failure, we improved the durability of blood chamber and evaluated the performance of the modified system via the hemolysis test.
Keywords:Extracorporeal life support systems (ECLS)  Reliability  Durability  Blood chamber  Hemolysis testing
本文献已被 PubMed SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号