Implantable cardiac pacemaker failure by cumulative dose effects of flattening filter free beams
Authors:Kazuhiko Nakamura  Takahiro Aoyama  Naoki Kaneda  Masashi Otsuji  Yoshitaka Minami  Ami Sakuragi  Masaru Nakamura
Affiliation:Department of Radiology, Aichi Medical University Hospital, 1-1 Yazakokarimata, Nagakute, Aichi 480-1195 Japan;Department of Radiation Oncology, Aichi Cancer Center, 1-1 Kanokoden, Chikusa-Ku, Nagoya, Aichi 464-8681 Japan;Graduate School of Medicine, Aichi Medical University, 1-1 Yazako-karimata, Nagakute, Aichi 480-1195 Japan;Department of Clinical Engineering, Aichi Medical University Hospital, 1-1 Yazakokarimata, Nagakute, Aichi 480-1195 Japan
Abstract:Cumulative dose effects, which are one of the main causes of errors that occur when an implantable cardiac pacemaker (ICP) is irradiated with ionizing radiation, induce permanent failure in ICPs. Although flattening filter free (FFF) beams, which are often used in stereotactic radiotherapy, are known to have different characteristics from conventional (with flattening filter [WFF]) beams, the cumulative dose effects on ICPs with FFF beams have been under-investigated. This study investigates ICP failure induced by cumulative dose effects of FFF beams. When the ICP placed in the center of the irradiation field was irradiated with 10 MV-FFF at 24 Gy/min, the cumulative dose at which failure occurred was evaluated on the basis of the failure criteria associated with high cumulative dose as described in the American Association of Physicists in Medicine Task Group 203. The ICP failures such as a mild battery depletion at a cumulative dose of 10 Gy, pacing-output voltage change >25% at a cumulative dose of 122 Gy, and the loss of telemetry capability at cumulative dose 134 Gy were induced by cumulative dose effects. The cumulative doses by which the cumulative dose effects of FFF beams induced ICP failure were not very different from those reported in previous studies with WFF beams. Therefore, radiotherapy with FFF beams (and WFF beams) for patients with ICP requires appropriate management for minimizing the cumulative dose effects.
Keywords:FFF beams   implantable cardiac pacemaker (ICP)   cumulative dose effects   failure   radiotherapy
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号